Endotracheal tube placement confirmation: 100% sensitivity and specificity with sustained four-phase capnographic waveforms in a cadaveric experimental model.
Resuscitation 2017 June
BACKGROUND: Waveform capnography is considered the gold standard for verification of proper endotracheal tube placement, but current guidelines caution that it is unreliable in low-perfusion states such as cardiac arrest. Recent case reports found that long-deceased cadavers can produce capnographic waveforms. The purpose of this study was to determine the predictive value of waveform capnography for endotracheal tube placement verification and detection of misplacement using a cadaveric experimental model.
METHODS: We conducted a controlled experiment with two intubated cadavers. Tubes were placed within the trachea, esophagus, and hypopharynx utilizing video laryngoscopy. We recorded observations of capnographic waveforms and quantitative end-tidal carbon dioxide (ETCO2 ) values during tracheal versus extratracheal (i.e., esophageal and hypopharyngeal) ventilations.
RESULTS: 106 and 89 tracheal ventilations delivered to cadavers one and two, respectively (n=195) all produced characteristic alveolar waveforms (positive) with ETCO2 values ranging 2-113mmHg. 42 esophageal ventilations (36 to cadaver one and 6 to cadaver two), and 6 hypopharyngeal ventilations (4 to cadaver one and 2 to cadaver two) all resulted in non-alveolar waveforms (negative) with ETCO2 values of 0mmHg. Esophageal and hypopharyngeal measurements were categorized as extratracheal (n=48). A binary classification test showed no false negatives or false positives, indicating 100% sensitivity (NPV 1.0, 95%CI 0.98-1.00) and 100% specificity (PPV 1.0, 95%CI 0.93-1.00).
CONCLUSION: Though current guidelines question the reliability of waveform capnography for verifying endotracheal tube location during low-perfusion states such as cardiac arrest, our findings suggest that it is highly sensitive and specific.
METHODS: We conducted a controlled experiment with two intubated cadavers. Tubes were placed within the trachea, esophagus, and hypopharynx utilizing video laryngoscopy. We recorded observations of capnographic waveforms and quantitative end-tidal carbon dioxide (ETCO2 ) values during tracheal versus extratracheal (i.e., esophageal and hypopharyngeal) ventilations.
RESULTS: 106 and 89 tracheal ventilations delivered to cadavers one and two, respectively (n=195) all produced characteristic alveolar waveforms (positive) with ETCO2 values ranging 2-113mmHg. 42 esophageal ventilations (36 to cadaver one and 6 to cadaver two), and 6 hypopharyngeal ventilations (4 to cadaver one and 2 to cadaver two) all resulted in non-alveolar waveforms (negative) with ETCO2 values of 0mmHg. Esophageal and hypopharyngeal measurements were categorized as extratracheal (n=48). A binary classification test showed no false negatives or false positives, indicating 100% sensitivity (NPV 1.0, 95%CI 0.98-1.00) and 100% specificity (PPV 1.0, 95%CI 0.93-1.00).
CONCLUSION: Though current guidelines question the reliability of waveform capnography for verifying endotracheal tube location during low-perfusion states such as cardiac arrest, our findings suggest that it is highly sensitive and specific.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app