Suitability of the DRI Hydrocodone/Hydromorphone Immunoassay in the Clinical Environment at a Lower Cutoff: Validation With LC-MS/MS Analysis

Robert B Dixon, Amitava Dasgupta
Therapeutic Drug Monitoring 2016, 38 (6): 787-790
27764026

BACKGROUND: We evaluated the analytical performance of the DRI hydrocodone/hydromorphone assay by comparing semiquantitative values obtained by this assay with values obtained by a liquid chromatography combined with tandem mass spectrometry (LC-MS/MS) method. We also evaluated the possibility of lowering the cutoff of the DRI assay from 300 to 100 ng/mL.

METHODS: We compared semiquantitative values obtained by the DRI assay in 97 specimens with values obtained by the LC-MS/MS method including 10 specimens containing hydrocodone and/or hydromorphone concentrations between 105.0 and 145.0 ng/mL (determined by LC-MS/MS) to determine the sensitivity at 100 ng/mL. In addition, several opioids at a concentration of 5000 ng/mL were also analyzed by the DRI assay to determine its specificity.

RESULTS: We observed no false-negative result using the DRI immunoassay in 96 specimens that showed semiquantitative values at 100 ng/mL or higher. However, one specimen containing 110 ng/mL of hydrocodone was false negative with the DRI assay (semiquantitative value 88 ng/mL, below 100 ng/mL cutoff). The semiquantitative values produced by DRI showed poor correlation with values determined by the LC-MS/MS method. The sensitivity of the DRI assay at 100 ng/mL was 90%, and the assay was very specific showing minimal cross-reactivity only with oxycodone and oxymorphone.

CONCLUSIONS: DRI immunoassay for hydrocodone/hydromorphone is a cost-effective method of screening urine specimens in the clinical environment at a lower cutoff of 100 ng/mL.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Trending Papers

Remove bar
Read by QxMD icon Read
27764026
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"