Evaluation of the stiffness characteristics of rapid palatal expander screws.
Progress in Orthodontics 2016 December
BACKGROUND: The aim of this study is to evaluate the mechanical properties of the screws used for rapid expansion of the upper jaw.
METHODS: Ten types of expansion screw were assessed, seven with four arms: Lancer Philosophy 1, Dentaurum Hyrax Click Medium, Forestadent Anatomic Expander type "S", Forestadent Anatomic Expander type "S" for narrow palates, Forestadent Memory, Leone A 2620-10 with telescopic guide, and Leone A 0630-10 with orthogonal arms; and three with two arms: Dentaurum Variety S.P., Target Baby REP Veltri, and Leone A 362113. A test expander with the mean dimensions taken from measurements on a sample of 100 expanders was constructed for each screw. The test expanders were connected to the supports of an Instron 4467 (Instron Corp., USA) mechanical testing machine equipped with a 500 N load cell, and the compression force exerted after each activation was measured. The mean forces expressed by the two- and four-arm expanders were then compared.
RESULTS: After five activations, the forces expressed by the two-arm devices were double than those expressed by the four-arm devices on average (224 ± 59.9 N vs. 103 ± 32.9 N), and such values remained high after subsequent activations.
CONCLUSIONS: The expanders tested demonstrated stiffness characteristics compatible with opening of the palatine sutures in pre-adolescent patients. The stiffness of such devices can be further increased during the construction phase.
METHODS: Ten types of expansion screw were assessed, seven with four arms: Lancer Philosophy 1, Dentaurum Hyrax Click Medium, Forestadent Anatomic Expander type "S", Forestadent Anatomic Expander type "S" for narrow palates, Forestadent Memory, Leone A 2620-10 with telescopic guide, and Leone A 0630-10 with orthogonal arms; and three with two arms: Dentaurum Variety S.P., Target Baby REP Veltri, and Leone A 362113. A test expander with the mean dimensions taken from measurements on a sample of 100 expanders was constructed for each screw. The test expanders were connected to the supports of an Instron 4467 (Instron Corp., USA) mechanical testing machine equipped with a 500 N load cell, and the compression force exerted after each activation was measured. The mean forces expressed by the two- and four-arm expanders were then compared.
RESULTS: After five activations, the forces expressed by the two-arm devices were double than those expressed by the four-arm devices on average (224 ± 59.9 N vs. 103 ± 32.9 N), and such values remained high after subsequent activations.
CONCLUSIONS: The expanders tested demonstrated stiffness characteristics compatible with opening of the palatine sutures in pre-adolescent patients. The stiffness of such devices can be further increased during the construction phase.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app