Impact of suture mediated femoral access site closure with the Prostar XL compared to the ProGlide system on outcome in transfemoral aortic valve implantation

Julia Seeger, Birgid Gonska, Christoph Rodewald, Wolfgang Rottbauer, Jochen Wöhrle
International Journal of Cardiology 2016 November 15, 223: 564-567
27561160

BACKGROUND: Management of femoral access site is an important issue in transfemoral transcatheter aortic valve implantation (TAVI) and crucial for acute and long-term outcome. Data on vascular closure devices in this setting are limited. We evaluated safety and efficacy of the Prostar XL compared to the ProGlide suture-based vascular closure device.

METHODS AND RESULTS: We enrolled 585 patients undergoing percutaneous transfemoral transcatheter aortic valve implantation (TAVI). Outcomes were defined according to Valve academic research consortium (VARC)-2 criteria. In 237 (40.5%) patients femoral access site closure was performed using the Prostar and in 348 patients (59.6%) using the ProGlide vascular closure device. There was no significant difference in patient baseline characteristics including single and dual antiplatelet therapies. Sheath outer diameter was significantly larger in the ProGlide compared with the Prostar group (7.7±1.5 vs. 7.9±0.5mm; p=0.001). Closure device failure according to VARC-2 criteria was significantly more frequent with the Prostar versus ProGlide device (19% vs. 4.6%; p<0.01). Need for surgical repair (11.8% vs. 0%, p<0.01), major (12.2% vs. 2.3%, p<0.01) and minor (17.3% vs. 5.7%, p<0.01) vascular complications and bleeding complications (5.5% vs. 2.0%, p=0.02) occurred significantly more often with the Prostar device compared with the ProGlide system. In addition, in-hospital mortality was higher with Prostar compared with ProGlide (5.9% vs. 2.0%; p=0.01).

CONCLUSION: Femoral access site closure with the ProGlide device compared with the Prostar device in transfemoral aortic valve implantation was associated with significantly lower rates of closure device failure, minor and major bleedings and a significantly lower in-hospital mortality.

CLINICAL TRIAL REGISTRATION: clinicaltrials.govNCT02162069.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
27561160
×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"