The Use of Cone Beam Computed Tomography for Image Guided Gamma Knife Stereotactic Radiosurgery: Initial Clinical Evaluation.

PURPOSE: The present study used cone beam computed tomography (CBCT) to measure the inter- and intrafraction uncertainties for intracranial stereotactic radiosurgery (SRS) using the Leksell Gamma Knife (GK).

METHODS AND MATERIALS: Using a novel CBCT system adapted to the GK radiosurgery treatment unit, CBCT images were acquired immediately before and after treatment for each treatment session within the context of a research ethics board-approved prospective clinical trial. Patients were immobilized in the Leksell coordinate frame (LCF) for both volumetric CBCT imaging and GK-SRS delivery. The relative displacement of the patient's skull to the stereotactic reference (interfraction motion) was measured for each CBCT scan. Differences between the pre- and post-treatment CBCT scans were used to determine the intrafraction motion.

RESULTS: We analyzed 20 pre- and 17 post-treatment CBCT scans in 20 LCF patients treated with SRS. The mean translational pretreatment setup error ± standard deviation in the left-right, anteroposterior, and craniocaudal directions was -0.19 ± 0.32, 0.06 ± 0.27, and -0.23 ± 0.2 mm, with a maximum of -0.74, -0.53, and -0.68 mm, respectively. After an average time between the pre- and post-treatment CBCT scans of 82 minutes (range 27-170), the mean intrafraction error ± standard deviation for the LCF was -0.03 ± 0.05, -0.03 ± 0.18, and -0.03 ± 0.12 mm in the left-right, anteroposterior, and craniocaudual direction, respectively.

CONCLUSIONS: Using CBCT on a prototype image guided GK Perfexion unit, we were able to measure the inter- and intrafraction positional changes for GK-SRS using the invasive frame. In the era of image guided radiation therapy, the use of CBCT image guidance for both frame- and non-frame-based immobilization systems could serve as a useful quality assurance tool. Our preliminary measurements can guide the application of achievable thresholds for inter- and intrafraction discrepancy when moving to a frameless approach.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app