Reducing bipolar sterilization failures.
Obstetrics and Gynecology 1989 July
Many bipolar sterilization failures have occurred because of incomplete desiccation of the endosalpinx. This study compared the effect upon human fallopian tubes of different waveforms of electrical energy at advancing power settings. When the Valleylab SSE2-L generator was attached to the Kleppinger bipolar forceps, complete desiccation was confirmed with the power set at 25 W in a cutting waveform. Other waveforms (coagulation and blend) failed to complete the task at the same power setting. An inline current meter assures the operator that all the available energy is delivered. This information should help to reduce bipolar sterilization failures.
Full text links
Trending Papers
How to improve the efficiency and the safety of real-time ultrasound-guided central venous catheterization in 2023: a narrative review.Annals of Intensive Care 2023 May 26
SGLT2 Inhibitors: A New Therapeutical Strategy to Improve Clinical Outcomes in Patients with Chronic Kidney Diseases.International Journal of Molecular Sciences 2023 May 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app