OBJECTIVES: Our objectives were to assess for differences between standard and radiofrequency (RF) needle in procedural times, success, and spatial accuracy.

BACKGROUND: Targeted transseptal (TS) puncture is essential for structural heart interventions. Spatial accuracy of standard versus RF needle has not been reported.

METHODS: Consecutive patients undergoing structural heart interventions requiring TS puncture were studied retrospectively. A standard needle was alternated with a RF needle. Procedural success and times were recorded. Measurements based on intraprocedural transesophageal echocardiograms of the intended versus final TS crossing site were obtained. Pre-puncture and maximal tenting of the septum were also recorded.

RESULTS: Twenty-five patients underwent standard needle and 27 RF TS access. All RF needle attempts to cross were successful without an assistance wire. Three standard needle patients required assisted crossing. After failed TS access, two patients had successful RF TS access. TS procedural times significantly favored the RF needle for time from septum to puncture (P = 0.02). Both standard and RF access yielded accurate crossing with no significant differences between the intended and actual crossing site. Maximal tenting was significantly less with the RF needle (P = 0.004). There were no major complications.

CONCLUSIONS: In the hands of an experienced operator, there were no major clinically significant differences between the standard and RF needle approaches. However, procedural unassisted crossing success was higher with the RF needle approach as compared to standard needle. Procedural times and degree of septum tenting favored the RF over standard needle. Accuracy was similar with both approaches. RF was a successful strategy when standard needle failed. There were no major complications with either TS puncture strategy. © 2016 Wiley Periodicals, Inc.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
27220107
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"

We want to hear from doctors like you!

Take a second to answer a survey question.