Insulation failure in electrosurgery instrumentation: a prospective evaluation.
Surgical Endoscopy 2016 November
BACKGROUND: The use of electrosurgery has expanded to a wide variety of surgical specialities, but it has also been accompanied by its share of complications, including thermal injuries to nontargeted tissues, caused by a break or defect in the insulation of the instrument's coat. The purpose of this study was to determine the prevalence and the location of insulation failures (IFs) in electrosurgical instruments, then to assess the necessity of routine IF testing.
METHODS: Electrosurgical instruments were visually inspected and checked for IF using a high-voltage detector. Two different detectors were used during two testing sessions: DTU-6 (Petel company) and DIATEG (Morgate company). Laparoscopic and non-laparoscopic instruments were determined to have IF if current crossed the instrument's insulation, signaled by an alarm sound.
RESULTS: A total of 489 instruments were tested. The overall prevalence of IFs was 24.1 % with only visual inspection and 37.2 % with the IF detector. Among the 489 instruments, 13.1 % were visually intact, but had an electric test failure. DTU-6 and DIATEG detectors showed comparable efficiency in detection of overall IFs and for laparoscopic and non-laparoscopic instruments. The median location of IFs was more pronounced for laparoscopic instruments (50.4 %) and the distal location for non-laparoscopic instruments (40.4 %).
CONCLUSION: Accidental burns are a hidden problem and can lead to patient complications. In Central Sterilization Service Department, prevention currently includes only visual control of electrosurgery instrumentation, but testing campaigns are now necessary in order to identify maximum instruments' defects.
METHODS: Electrosurgical instruments were visually inspected and checked for IF using a high-voltage detector. Two different detectors were used during two testing sessions: DTU-6 (Petel company) and DIATEG (Morgate company). Laparoscopic and non-laparoscopic instruments were determined to have IF if current crossed the instrument's insulation, signaled by an alarm sound.
RESULTS: A total of 489 instruments were tested. The overall prevalence of IFs was 24.1 % with only visual inspection and 37.2 % with the IF detector. Among the 489 instruments, 13.1 % were visually intact, but had an electric test failure. DTU-6 and DIATEG detectors showed comparable efficiency in detection of overall IFs and for laparoscopic and non-laparoscopic instruments. The median location of IFs was more pronounced for laparoscopic instruments (50.4 %) and the distal location for non-laparoscopic instruments (40.4 %).
CONCLUSION: Accidental burns are a hidden problem and can lead to patient complications. In Central Sterilization Service Department, prevention currently includes only visual control of electrosurgery instrumentation, but testing campaigns are now necessary in order to identify maximum instruments' defects.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
Perioperative echocardiographic strain analysis: what anesthesiologists should know.Canadian Journal of Anaesthesia 2024 April 11
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app