Inter- and intra-fraction motion in stereotactic body radiotherapy for spinal and paraspinal tumours using cone-beam CT and positional correction in six degrees of freedom

Renee Finnigan, Brock Lamprecht, Tamara Barry, Kimberley Jones, Joshua Boyd, Andrew Pullar, Bryan Burmeister, Matthew Foote
Journal of Medical Imaging and Radiation Oncology 2016, 60 (1): 112-8
26813044

INTRODUCTION: Stereotactic body radiotherapy (SBRT) for spinal tumours delivers high doses per fraction to targets in close proximity to neural tissue. With steep dose gradients, small changes in position can confer significant dosimetric impact on adjacent structures. We analysed positioning error in consecutively treated patients on a strict image-guidance protocol with online correction in 6 degrees of freedom (6-DOF).

METHODS: Set-up error, residual error post-correction and intra-fraction motion for 30 courses of spinal SBRT in 27 patients were assessed using cone-beam CT. Positional error was corrected in x, y and z translational planes and rotational axes using a robotic couch, applying 2 mm and 2° action levels. Linear mixed-effects model assessed whether positional error was influenced by factors such as vertebral level, immobilisation device and treatment duration.

RESULTS: Sixty-two fractions were delivered with 225 image registrations. Median treatment duration was significantly longer for patients treated with static-field intensity-modulated radiotherapy compared with volumetric-modulated arc treatment--40 min versus 28 min, respectively (P = 0.01). Across all fractions, the median residual positional error after initial correction was greatest in the x translational plane (0.5 mm; 95% confidence interval (CI) 0.3-0.6) and y rotational axis (0.25°; 95% CI 0.1-0.3). Median intra-fraction error was also greatest in the x-plane (0.7 mm; 95% CI 0.5-1.0) and y-axis (0.4°; 95% CI 0.2-0.5).

CONCLUSION: With strict immobilisation, image-guidance and 6-DOF correction, our current practice of applying 3-mm planning margins for target volumes and critical structures appears safe. Lower image-guidance action thresholds plus verification with end-to-end testing would be recommended before further reducing margins.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
26813044
×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"