JOURNAL ARTICLE
RESEARCH SUPPORT, NON-U.S. GOV'T
Add like
Add dislike
Add to saved papers

MRI-Based Attenuation Correction for PET/MRI Using Multiphase Level-Set Method.

UNLABELLED: Inaccuracy in MR image-based attenuation correction (MR-AC) leads to errors in quantification and the misinterpretation of lesions in brain PET/MRI studies. To resolve this problem, we proposed an improved ultrashort echo time MR-AC method that was based on a multiphase level-set algorithm with main magnetic field (B0) inhomogeneity correction. We also assessed the feasibility of this level-set-based MR-AC method (MR-AC(level)), compared with CT-AC and MR-AC provided by the manufacturer of the PET/MRI scanner (MR-AC(mMR)).

METHODS: Ten healthy volunteers and 20 Parkinson disease patients underwent(18)F-FDG and(18)F-fluorinated-N-3-fluoropropyl-2-β-carboxymethoxy-3-β-(4-iodophenyl)nortropane ((18)F-FP-CIT) PET scans, respectively, using both PET/MRI and PET/CT scanners. The level-set-based segmentation algorithm automatically delimited air, bone, and soft tissue from the ultrashort echo time MR images. For the comparison, MR-AC maps were coregistered to reference CT. PET sinogram data obtained from PET/CT studies were then reconstructed using the CT-AC, MR-AC(mMR), and MR-AC(level) maps. The accuracies of SUV, SUVr (SUV and its ratio to the cerebellum), and specific-to-nonspecific binding ratios obtained using MR-AC(level) and MR-AC(mMR) were compared with CT-AC using region-of-interest- and voxel-based analyses.

RESULTS: There was remarkable improvement in the segmentation of air cavities and bones and the quantitative accuracy of PET measurement using the level set. Although the striatal and cerebellar activities in (18)F-FP-CIT PET and frontal activity in (18)F-FDG PET were significantly underestimated by the MR-AC(mMR), the MR-AC(level) provided PET images almost equivalent to the CT-AC images. PET quantification error was reduced by a factor of 3 using MR-AC(level) (SUV error < 10% in MR-AC(level) and < 30% in MR-AC(mMR) [version VB18P], and < 5% in MR-AC(level) and < 15% in MR-AC(mMR) [VB20P]).

CONCLUSION: The results of this study indicate that our new multiphase level-set-based MR-AC method improves the quantitative accuracy of brain PET in PET/MRI studies.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app