Feasibility and safety of adenosine cardiovascular magnetic resonance in patients with MR conditional pacemaker systems at 1.5 Tesla.

BACKGROUND: Cardiovascular Magnetic Resonance (CMR) with adenosine stress is a valuable diagnostic tool in coronary artery disease (CAD). However, despite the development of MR conditional pacemakers CMR is not yet established in clinical routine for pacemaker patients with known or suspected CAD. A possible reason is that adenosine stress perfusion for ischemia detection in CMR has not been studied in patients with cardiac conduction disease requiring pacemaker therapy. Other than under resting conditions it is unclear whether MR safe pacing modes (paused pacing or asynchronous mode) can be applied safely because the effect of adenosine on heart rate is not precisely known in this entity of patients. We investigate for the first time feasibility and safety of adenosine stress CMR in pacemaker patients in clinical routine and evaluate a pacing protocol that considers heart rate changes under adenosine.

METHODS: We retrospectively analyzed CMR scans of 24 consecutive patients with MR conditional pacemakers (mean age 72.1 ± 11.0 years) who underwent CMR in clinical routine for the evaluation of known or suspected CAD. MR protocol included cine imaging, adenosine stress perfusion and late gadolinium enhancement.

RESULTS: Pacemaker indications were sinus node dysfunction (n = 18) and second or third degree AV block (n = 6). Under a pacing protocol intended to avoid competitive pacing on the one hand and bradycardia due to AV block on the other no arrhythmia occurred. Pacemaker stimulation was paused to prevent competitive pacing in sinus node dysfunction with resting heart rate >45 bpm. Sympatho-excitatory effect of adenosine led to a significant acceleration of heart rate by 12.3 ± 8.3 bpm (p < 0.001), no bradycardia occurred. On the contrary in AV block heart rate remained constant; asynchronous pacing above resting heart rate did not interfere with intrinsic rhythm.

CONCLUSION: Adenosine stress CMR appears to be feasible and safe in patients with MR conditional pacemakers. Heart rate response to adenosine has to be considered for the choice of pacing modes during CMR.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app