Comparative Study
Journal Article
Add like
Add dislike
Add to saved papers

High Risk of Failure With Bimodular Femoral Components in THA.

BACKGROUND: The bimodular femoral neck implant (modularity in the neck section and prosthetic head) offers several implant advantages to the surgeon performing THAs, however, there have been reports of failure of bimodular femoral implants involving neck fractures or adverse tissue reaction to metal debris. We aimed to assess the results of the bimodular implants used in the THAs we performed.

QUESTIONS/PURPOSES: We asked: (1) What is the survivorship of the PROFEMUR(®) bimodular femoral neck stems? (2) What are the modes of failure of this bimodular femoral neck implant? (3) What are the major risk factors for the major modes of failure of this device?

METHODS: Between 2003 and 2009, we used one family of bimodular femoral neck stems for all primary THAs (PROFEMUR(®) Z and PROFEMUR(®) E). During this period, 277 THAs (in 242 patients) were performed with these implants. One hundred seventy were done with the bimodular PROFEMUR(®) E (all are accounted for here), and when that implant was suspected of having a high risk of failure, the bimodular PROFEMUR(®) Z was used instead. One hundred seven THAs were performed using this implant (all are accounted for in this study). All bearing combinations, including metal-on-metal, metal-on-polyethylene, and ceramic-on-ceramic, are included here. Data for the cohort included patient demographics, BMI, implant dimensions, type of articular surface, length of followup, and C-reactive protein serum level. We assessed survivorship of the two stems using Kaplan-Meier curves and determined the frequency of the different modes of stem failure. For each of the major modes of failure, we performed binary logistic regression to identify associated risk factors.

RESULTS: Survivorship of the stems, using aseptic revision as the endpoint, was 85% for the patients with the PROFEMUR(®) E stems with a mean followup of 50 months (range, 1-125 months) and 85% for the PROFEMUR(®) Z with a mean followup of 50 months (range, 1-125 months)(95% CI, 74-87 months). The most common modes of failure were loosening (9% for the PROFEMUR(®) E), neck fracture (6% for the PROFEMUR(®) Z and 0.6% for the PROFEMUR(®) E), metallosis (1%), and periprosthetic fracture (1%). Only the bimodular PROFEMUR(®) E was associated with femoral stem loosening (odds ratio [OR] =1.1; 95% CI, 1.04-1.140; p = 0.032). Larger head (OR = 3.2; 95% CI, 0.7-14; p = 0.096), BMI (OR = 1.19; 95% CI, 1-1.4; p = 0.038) and total offset (OR = 1.83; 95% CI, 1.13-2.9; p = 0.039) were associated with neck fracture.

CONCLUSION: Bimodular neck junctions may be potentiated by long neck lengths, greater offset, and larger head diameters. These factors may contribute to bimodular neck failure by creating a larger moment about the neck's insertion in the stem. The PROFEMUR(®) E implant is associated with high periprosthetic loosening. Based on our experience we cannot recommend the use of bimodular femoral neck implants.

LEVEL OF EVIDENCE: Level III, therapeutic study.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app