EVALUATION STUDIES
JOURNAL ARTICLE
Add like
Add dislike
Add to saved papers

The impact of transmission-emission misregistration on the interpretation of SPET/CT myocardial perfusion studies and the value of misregistration correction.

OBJECTIVE: Previous studies indicate that the quality of single photon emission tomography/computed tomography (SPET/CT) myocardial perfusion imaging (MPI) is degraded by even mild transmission-emission misregistrations. The purpose of the current study was to investigate the impact of SPET/CT misalignment on the interpretation of MPI and examine the value of a commercial software application for registration correction.

SUBJECTS AND METHODS: A total of 255 technetium-99m ((99m)Tc)-tetrofosmin stress/rest MPI examinations in 150 patients were reviewed for SPET/CT misalignment. After registration correction by the software, images were reassessed for interpretation differences from the misregistered study. The diagnostic benefit of reregistration was determined by taking into account the non-attenuation compensated image pattern, combined stress-rest evaluation, gated-SPET data and patient's history. In a phantom experiment and in 3 representative clinical cases, SPET/CT misalignment was purposely created by the software by sequential slice shifts and its effect was evaluated quantitatively.

RESULTS: Misregistration ≥1 pixel in at least one direction was observed in 24% of studies. Interpretation of MPI changed after registration correction in 11% of cases with misalignment <1 pixel, in 18% with 1-2 and in 73% with ≥2 pixels. The diagnostic information seemed to improve after registration correction in 58% of studies irrespective of the degree of misregistration. Software-simulated misregistration had dissimilar effects in the phantom and the 3 selected clinical cases.

CONCLUSIONS: The impact of SPET/CT misregistration on MPI interpretation although influenced by the degree and direction of slice misplacement, it is also case-specific and hardly predictable. Registration restoration by the software seems worthwhile regardless of misregistration magnitude.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app