Evaluation of the COBAS(®) AmpliPrep/COBAS(®) TaqMan(®) HCV Test, v2.0 and comparison to assays used in routine clinical practice in an international multicenter clinical trial: The ExPECT study.
Journal of Clinical Virology 2015 June
BACKGROUND: The COBAS(®) AmpliPrep(®)/COBAS(®) TaqMan(®) HCV Test, v2.0 (CAP/CTM2) is used for HCV RNA viral load monitoring.
OBJECTIVES: The performance of the CAP/CTM2 was compared to other widely used tests, including a manual version of the assay (the COBAS(®) TaqMan(®) HCV Test, v2.0 for use with the High Pure System, HPS/CTM2) predominantly used during phase III clinical trials for the new direct acting antiviral therapies.
STUDY DESIGN: Low HCV RNA level comparisons were performed across tests (Abbott Realtime HCV Test, ART; COBAS(®) AmpliPrep(®)/COBAS(®) TaqMan(®) HCV Test, v1.0, CAP/CTM1; CAP/CTM2; and HPS/CTM2) using dilutions of the 2nd HCV WHO International Standard. Additionally, the clinical performance of the CAP/CTM2 was evaluated with 421 leftover HCV RNA-positive routine clinical samples.
RESULTS: All quantifiable WHO dilutions were within ±0.3log10IU/mL of the expected results across tests and the analytical sensitivity resulted in a limit of detection of 12IU/mL (95% confidence interval, 10, 15). When clinical samples were tested the results for 87% (367 of 421) of all sample comparisons were within ±0.5log10IU/mL. When low viral load results (25-3500IU/mL) were compared, values obtained by the ART assay were significantly lower (p<0.0001) than those obtained with the CAP/CTM2.
CONCLUSIONS: The new CAP/CTM2 showed good accuracy with comparable sensitivity to comparator assays. The new kit is well-suited for use in the routine diagnostic laboratory, especially for accurate monitoring of patients receiving triple therapy or interferone-free regimens.
OBJECTIVES: The performance of the CAP/CTM2 was compared to other widely used tests, including a manual version of the assay (the COBAS(®) TaqMan(®) HCV Test, v2.0 for use with the High Pure System, HPS/CTM2) predominantly used during phase III clinical trials for the new direct acting antiviral therapies.
STUDY DESIGN: Low HCV RNA level comparisons were performed across tests (Abbott Realtime HCV Test, ART; COBAS(®) AmpliPrep(®)/COBAS(®) TaqMan(®) HCV Test, v1.0, CAP/CTM1; CAP/CTM2; and HPS/CTM2) using dilutions of the 2nd HCV WHO International Standard. Additionally, the clinical performance of the CAP/CTM2 was evaluated with 421 leftover HCV RNA-positive routine clinical samples.
RESULTS: All quantifiable WHO dilutions were within ±0.3log10IU/mL of the expected results across tests and the analytical sensitivity resulted in a limit of detection of 12IU/mL (95% confidence interval, 10, 15). When clinical samples were tested the results for 87% (367 of 421) of all sample comparisons were within ±0.5log10IU/mL. When low viral load results (25-3500IU/mL) were compared, values obtained by the ART assay were significantly lower (p<0.0001) than those obtained with the CAP/CTM2.
CONCLUSIONS: The new CAP/CTM2 showed good accuracy with comparable sensitivity to comparator assays. The new kit is well-suited for use in the routine diagnostic laboratory, especially for accurate monitoring of patients receiving triple therapy or interferone-free regimens.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app