A quality improvement initiative to optimize use of a mechanical chest compression device within a high-performance CPR approach to out-of-hospital cardiac arrest resuscitation.

Resuscitation 2015 July
BACKGROUND: Minimizing the chest compression pause associated with application of a mechanical CPR device is a key component of optimal integration into the overall resuscitation process. As part of a multi-agency implementation project, Anchorage Fire Department deployed LUCAS CPR devices on BLS and ALS fire apparatus for initiation early in resuscitation efforts. A 2012 report identified the pause interval for device application as a key opportunity for quality improvement (QI). In early 2013 we began a QI initiative to reduce device application time interval and optimize the overall CPR process. To assess QI initiative effectiveness, we compared key CPR process metrics from before to during and after its implementation.

METHODS: We included all cases of EMS-treated out-of-hospital cardiac arrest during 2012 and 2013 in which a mechanical CPR device was used and the defibrillator electronic record was available. Continuous ECG and impedance data were analyzed to measure chest compression fraction, duration of the pause from last manual to first mechanical compression, and duration of the longest overall pause in the resuscitation effort.

RESULTS: Compared to cases from 2012 (n = 61), median duration of the pause prior to first mechanical compression for cases from 2013 (n = 71) decreased from 21 (15, 31) to 7 (4, 12)s (p < 0.001), while median chest compression fraction increased from 0.90 (0.88, 0.93) to 0.95 (0.93, 0.96) (p < 0.001). Median duration of the longest pause decreased from 25 (20, 35) to 13 (10, 20)s (p < 0.001), while the proportion of cases where the longest pause was for mechanical CPR application decreased from 74% to 31% (p < 0.001).

CONCLUSIONS: Our QI initiative substantially reduced the duration of the pause prior to first mechanical compression. Combined with the simultaneous significant increase in compression fraction and significant decrease in duration of the longest pause, this finding strongly suggests a large improvement in mechanical CPR device application efficiency within an overall high-performance CPR process.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app