Anti-factor Xa and activated partial thromboplastin time measurements for heparin monitoring in mechanical circulatory support.

OBJECTIVES: This study investigated the relationship between anti-factor Xa (anti-FXa) and activated partial thromboplastin time (aPTT) for monitoring intravenous unfractionated heparin (IV-UFH) in patients with continuous-flow left ventricular assist devices (CF-LVADs).

BACKGROUND: CF-LVADs have become mainstream therapy for patients with advanced heart failure. Thromboembolic events, device thrombosis, and bleeding continue to be a challenge with this technology. Adequate anticoagulation is required to prevent these adverse events.

METHODS: A prospective study of consecutive patients implanted with a CF-LVAD was conducted. Paired samples were considered concordant if aPTT values fell into expected ranges for subtherapeutic, therapeutic, and supratherapeutic anti-FXa levels. Heparin dosing was on the basis of anti-Xa levels.

RESULTS: A total of 340 paired values from 38 patients were evaluated. Anti-FXa and aPTT were discordant in 253 samples (74.4%), with a high degree of variability in aPTT for any given anti-FXa level (r(2) = 0.57). Results were discordant in 104 samples (63.8%) from patients undergoing bridging therapy with warfarin and in 149 samples (84.2%) from patients with device obstruction and/or hemolysis (p < 0.001). The most common pattern of discordance was a supratherapeutic aPTT value despite a therapeutic anti-FXa level (49.1% for bridging vs. 75.8% for device obstruction and/or hemolysis; p < 0.001).

CONCLUSIONS: Levels of aPTT were disproportionately prolonged relative to the corresponding anti-FXa levels in CF-LVAD patients, particularly those with device obstruction. Hemolysis and warfarin administration may falsely elevate aPTT, resulting in overestimation of heparin concentration and under-anticoagulation. Use of aPTT and anti-FXa to guide heparin therapy may lead to different estimates of heparin concentration in the same patient.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app