[Biomechanical testing of a new total wrist arthrodesis plate]

Stefan Passin, N Pochlatko, D Ebi, A Spiegel, M Sauerbier
Handchirurgie, Mikrochirurgie, Plastische Chirurgie 2015, 47 (1): 32-7
25706177

PURPOSE: The aim of this study was to evaluate the biomechanical properties of the APTUS wrist fusion plate in comparison to those of the AO wrist fusion plate.

MATERIAL AND METHOD: 6 APTUS wrist fusion plates (APTUS 2.5 TriLock Wrist Fusion Plate Long Bend) from Medartis (Basel, Switzerland) and 6 AO wrist fusion plates (= LCP Wrist Fusion Plate Standard Bend from Dupuy-Synthes, Bettlach, Switzerland) were installed according to the manufacturers' instructions on life-sized, fibre-glass reinforced polyamide models of a healthy human hand. Then fatigue testing with increasing loads on a monoaxial material testing machine was carried out and the force-movement measurements were recorded on a PC. For up to 50,000 cycles, loading was applied in a sinus-wave frequency of 4 Hz and a force ratio (FMin/FMax) of 0.1-70 N. After 50,000 cycles and then at every 10,000 cycles the load was increased by 15% until breakage or a deformation of over 15 mm at the point of force application of the implant occurred. In addition, the tear-out resistance of 5 screws of each type of plate was examined and the system tear-out resistance calculated. The mean values and standard deviations of fatigue strength, torsional moments and fatigue limits (load cycles) were checked for normal distribution and finally the results were compared by means of non-parametric and parametric statistical tests.

RESULTS: Plate breakage occurred with the AO wrist fusion plate on average after 52,596 (SD±12,833) load cycles and with the APTUS wrist fusion plate after an average of 115,428 (SD±12,600) load cycles. For the AO wrist fusion plates, an average load of 6.3 (SD±0.8) Nm and for the APTUS plate one of 10 (SD±1.7) Nm led to failure. For the APTUS plate the screw tear-out resistance was calculated as 2,632 (SD±96) N while that for the AO plate was found to be 1,449 (SD±314) N.

CONCLUSION: In comparison to the AO wrist fusion plate the APTUS wrist fusion plate has significantly higher fatigue resistance and tear-out strength. On the basis of 100,000 load cycles until bone healing in the case a wrist arthrodesis, use of the APTUS plate should enable bone healing to occur before the implant fails.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
25706177
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"