Significant pressure differences between solid-state and water-perfused systems in lower esophageal sphincter measurement.
Surgical Endoscopy 2015 December
OBJECTIVE: High-resolution manometry of the esophagus has gained worldwide acceptance, using different solid-state catheters. Thus, normal values for lower esophageal sphincter (LES) resting pressure in suspected gastroesophageal reflux disease patients have been established using water-perfused manometry. These standard values are commonly applied using also solid-state techniques, although they have never been compared before. The aim of the study was to compare LES measurements obtained with water-perfused manometry with a solid-state technique.
METHODS: Thirty healthy subjects were studied twice on the same day: Technique 1: Station pull through using a water-perfused catheter with ports arranged at 0°, 90°, 180° and 270° which were averaged to give a mean LES pressure. Technique 2: Solid-state circumferential probe with a single station pull through. Data were collected using the same computer system and program. The LES pressures were randomly and blindly analyzed.
RESULTS: Twenty-seven subjects out of 30 were analyzed. Using the solid-state system, the mean LES pressure was higher (15.0 vs. 23.3 mmHg, p = 0.003) and 19 of 27 (70%) individual measurements were higher. Two subjects had a hypertensive LES by solid state (58.6 resp. 47.5 mmHg), while their pressures were normal with water-perfused manometry (21.0 resp. 23.4 mmHg). The distal esophageal pressures (mean of pressure at 3 and 8 cm above LES) were the same with the two techniques.
CONCLUSION: In normal control subjects, LES measurement using circumferential solid-state transducers yields higher pressures than standard water-perfused manometry. Which system yields the "true" resting pressure of the physiologic LES remains to be determined.
METHODS: Thirty healthy subjects were studied twice on the same day: Technique 1: Station pull through using a water-perfused catheter with ports arranged at 0°, 90°, 180° and 270° which were averaged to give a mean LES pressure. Technique 2: Solid-state circumferential probe with a single station pull through. Data were collected using the same computer system and program. The LES pressures were randomly and blindly analyzed.
RESULTS: Twenty-seven subjects out of 30 were analyzed. Using the solid-state system, the mean LES pressure was higher (15.0 vs. 23.3 mmHg, p = 0.003) and 19 of 27 (70%) individual measurements were higher. Two subjects had a hypertensive LES by solid state (58.6 resp. 47.5 mmHg), while their pressures were normal with water-perfused manometry (21.0 resp. 23.4 mmHg). The distal esophageal pressures (mean of pressure at 3 and 8 cm above LES) were the same with the two techniques.
CONCLUSION: In normal control subjects, LES measurement using circumferential solid-state transducers yields higher pressures than standard water-perfused manometry. Which system yields the "true" resting pressure of the physiologic LES remains to be determined.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app