Three-dimensional printing of intracardiac defects from three-dimensional echocardiographic images: feasibility and relative accuracy.

BACKGROUND: With the advent of three-dimensional (3D) printers and high-resolution cardiac imaging, rapid prototype constructions of congenital cardiac defects are now possible. Typically, source images for these models derive from higher resolution, cross-sectional cardiac imaging, such as cardiac magnetic resonance imaging or computed tomography. These imaging methods may involve intravenous contrast, sedation, and ionizing radiation. New echocardiographic transducers and advanced software and hardware have optimized 3D echocardiographic images for this purpose. Thus, the objectives of this study were to confirm the feasibility of creating cardiac models from 3D echocardiographic data and to assess accuracy by comparing 3D model measurements with conventional two-dimensional (2D) echocardiographic measurements of cardiac defects.

METHODS: Nine patients undergoing 3D echocardiography were identified (eight with ventricular septal defects, one with three periprosthetic aortic valve leaks). Raw echocardiographic image data were exported anonymously and converted to Digital Imaging and Communications in Medicine format. The image data were filtered for noise reduction, imported into segmentation software to create a 3D digital model, and printed. Measurements of the defects from the 3D model were compared with defect measurements from conventional 2D echocardiographic data. Meticulous care was taken to ensure identical measurement planes.

RESULTS: Long- and short-axis measurements of eight ventricular septal defects and three perivalvar leaks were obtained. Mean ± SD values for the 3D model measurements and conventional 2D echocardiographic measurements were 7.5 ± 6.3 and 7.1 ± 6.2 mm respectively (P = .20), indicating no significant differences between the standard 2D and 3D model measurements. The two groups were highly correlated, with a Pearson correlation coefficient of 0.988. The mean absolute error (2D - 3D) for each measurement was 0.4 ± 0.9 mm, indicating accuracy of the 3D model of <1 mm.

CONCLUSIONS: Three-dimensional printed models of echocardiographic data are technically feasible and may accurately reflect ventricular septal defect anatomy. Three-dimensional models derived from 3D echocardiographic data sets represent a new tool in procedural planning for children with congenital heart disease.

Full text links

For the best experience, use the Read mobile app

Group 7SearchHeart failure treatmentPapersTopicsCollectionsEffects of Sodium-Glucose Cotransporter 2 Inhibitors for the Treatment of Patients With Heart Failure Importance: Only 1 class of glucose-lowering agents-sodium-glucose cotransporter 2 (SGLT2) inhibitors-has been reported to decrease the risk of cardiovascular events primarily by reducingSeptember 1, 2017: JAMA CardiologyAssociations of albuminuria in patients with chronic heart failure: findings in the ALiskiren Observation of heart Failure Treatment study.CONCLUSIONS: Increased UACR is common in patients with heart failure, including non-diabetics. Urinary albumin creatininineJul, 2011: European Journal of Heart FailureRandomized Controlled TrialEffects of Liraglutide on Clinical Stability Among Patients With Advanced Heart Failure and Reduced Ejection Fraction: A Randomized Clinical Trial.Review

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app