OPEN IN READ APP
JOURNAL ARTICLE
MULTICENTER STUDY

Multicenter outpatient dinner/overnight reduction of hypoglycemia and increased time of glucose in target with a wearable artificial pancreas using modular model predictive control in adults with type 1 diabetes

S Del Favero, J Place, J Kropff, M Messori, P Keith-Hynes, R Visentin, M Monaro, S Galasso, F Boscari, C Toffanin, F Di Palma, G Lanzola, S Scarpellini, A Farret, B Kovatchev, A Avogaro, D Bruttomesso, L Magni, J H DeVries, C Cobelli, E Renard
Diabetes, Obesity & Metabolism 2015, 17 (5): 468-76
25600304

AIMS: To test in an outpatient setting the safety and efficacy of continuous subcutaneous insulin infusion (CSII) driven by a modular model predictive control (MMPC) algorithm informed by continuous glucose monitoring (CGM) measurement.

METHODS: 13 patients affected by type 1 diabetes participated to a non-randomized outpatient 42-h experiment that included two evening meals and overnight periods (in short, dinner & night periods). CSII was patient-driven during dinner & night period 1 and MMPC-driven during dinner&night period 2. The study was conducted in hotels, where patients could move around freely. A CGM system (G4 Platinum; Dexcom Inc., San Diego, CA, USA) and insulin pump (AccuChek Combo; Roche Diagnostics, Mannheim, Germany) were connected wirelessly to a smartphone-based platform (DiAs, Diabetes Assistant; University of Virginia, Charlottesville, VA, USA) during both periods.

RESULTS: A significantly lower percentage of time spent with glucose levels <3.9 mmol/l was achieved in period 2 compared with period 1: 1.96 ± 4.56% vs 12.76 ± 15.84% (mean ± standard deviation, p < 0.01), together with a greater percentage of time spent in the 3.9-10 mmol/l target range: 83.56 ± 14.02% vs 62.43 ± 29.03% (p = 0.04). In addition, restricting the analysis to the overnight phases, a lower percentage of time spent with glucose levels <3.9 mmol/l (1.92 ± 4.89% vs 12.7 ± 19.75%; p = 0.03) was combined with a greater percentage of time spent in 3.9-10 mmol/l target range in period 2 compared with period 1 (92.16 ± 8.03% vs 63.97 ± 2.73%; p = 0.01). Average glucose levels were similar during both periods.

CONCLUSIONS: The results suggest that MMPC managed by a wearable system is safe and effective during evening meal and overnight. Its sustained use during this period is currently being tested in an ongoing randomized 2-month study.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Available on the App Store

Available on the Play Store
Remove bar
Read by QxMD icon Read
25600304
×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"