Clinical Biomechanics 2015 Februrary
BACKGROUND: This investigation compares an interspinous fusion device with posterior pedicle screw system in a lateral lumbar interbody lumbar fusion.

METHODS: We biomechanically tested six cadaveric lumbar segments (L1-L2) under an axial preload of 50N and torque of 5Nm in flexion-extension, lateral bending and axial rotation directions. We quantified range of motion, neutral zone/elastic zone stiffness in the following conditions: intact, lateral discectomy, lateral cage, cage with interspinous fusion, and cage with pedicle screws.

FINDINGS: A complete lateral discectomy and annulectomy increased motion in all directions compared to all other conditions. The lateral cage reduced motion in lateral bending and flexion/extension with respect to the intact and discectomy conditions, but had minimal effect on extension stiffness. Posterior instrumentation reduced motion, excluding interspinous augmentation in axial rotation with respect to the cage condition. Interspinous fusion significantly increased flexion and extension stiffness, while pedicle screws increased flexion/extension and lateral bending stiffness, with respect to the cage condition. Both posterior augmentations performed equivalently throughout the tests except in lateral bending stiffness where pedicle screws were stiffer in the neutral zone.

INTERPRETATION: A lateral discectomy and annulectomy generates immediate instability. Stand-alone lateral cages restore a limited amount of immediate stability, but posterior supplemental fixation increases stability. Both augmentations are similar in a single level lateral fusion in-vitro model, but pedicle screws are more equipped for coronal stability. An interspinous fusion is a less invasive alternative than pedicle screws and is potentially a conservative option for various interbody cage scenarios.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app