A comparison of fiberoptical guided tracheal intubation via laryngeal mask and laryngeal tube.
Saudi Journal of Anaesthesia 2015 January
BACKGROUND: Fiberoptical assisted intubation via a placed laryngeal mask airway (LMA) has been described as save and easy procedure to manage a difficult airway. The laryngeal tube (LT) is a promising alternative to the LMA as supraglottic airway device. Fiberoptical assisted intubation via LT is possible, however considered more difficult. The aim of this study was to compare the fiberoptical assisted intubation via LT and LMA.
MATERIALS AND METHODS: A total of 22 anesthesiologists with different levels of experience participated in the study performed on an adult airway model. Primarily the supraglottic device was placed and correct position was confirmed by successful ventilation. A 5 mm internal diameter tracheal tube was loaded onto a flexible 3.6 mm fiberscope and the so prepared device was inserted into the proximal lumen of the LMA or the LT. The glottis was passed under visual control and the tube advanced into the trachea. After removal of the fiberscope, ventilation was examined clinically by inspection. Success rates, procedure time and observed complications of LMA versus LT were compared (U-test; P < 0.05).
RESULTS: Placement of the endotracheal tube was successful in all attempts using both the LMA and LT. There was no difference in the time needed for the placement procedure (33 [26-38] s LMA; 35 [32-38] s LT). Only minor technical complications were observed in both groups.
CONCLUSION: A fiberoptical assisted intubation via LT can be considered as a relevant alternative in advanced airway management.
MATERIALS AND METHODS: A total of 22 anesthesiologists with different levels of experience participated in the study performed on an adult airway model. Primarily the supraglottic device was placed and correct position was confirmed by successful ventilation. A 5 mm internal diameter tracheal tube was loaded onto a flexible 3.6 mm fiberscope and the so prepared device was inserted into the proximal lumen of the LMA or the LT. The glottis was passed under visual control and the tube advanced into the trachea. After removal of the fiberscope, ventilation was examined clinically by inspection. Success rates, procedure time and observed complications of LMA versus LT were compared (U-test; P < 0.05).
RESULTS: Placement of the endotracheal tube was successful in all attempts using both the LMA and LT. There was no difference in the time needed for the placement procedure (33 [26-38] s LMA; 35 [32-38] s LT). Only minor technical complications were observed in both groups.
CONCLUSION: A fiberoptical assisted intubation via LT can be considered as a relevant alternative in advanced airway management.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app