[The study of corneal epithelial change in keratoconic eyes with Fourier-domain optical coherence tomographic pachymetry].

OBJECTIVE: To compare the corneal epithelial thickness profile in keratoconic and normal eyes with advanced Fourier-domain optical coherence tomography (OCT) and to evaluate the diagnostic efficiency of pachymetry parameters.

METHODS: Cross-sectional observational study. Twenty-five consecutive keratoconus patients (34 eyes) and 41 normal volunteers (41 eyes) in Eye & ENT Hospital of Fudan University between June 1 and July 31, 2013 were enrolled.Observations of corneal epithelial thickness and corneal thickness profiles were taken from all subjects. The epithelial pachymetry parameters were superior, inferior, inferior-superior (I-S), minimum (Min), maximum (Max), Min-Max, and central. The corneal pachymetry parameters were inferotemporal-superonasal (IT-SN), I-S, Min, Min-median, Min-Max, and central.Statistical analysis methods included Chi-square test, independent two-sample t-test, and Wilcoxon rank sum test. The diagnostic efficiency of all parameters were analyzed by receiver operating characteristic (ROC) curves and Youden index.

RESULTS: The inferior, I-S, Min, Min-Max, and central epithelial thickness averages were (50.5 ± 4.8) µm, (-2.7 ± 7.5) µm, (35.9 ± 10.7) µm, (-33.6 ± 18.9) µm, and (50.1 ± 6.8) µm in keratoconic eyes.In normal eyes, they were (54.8 ± 2.5) µm, (0.9 ± 1.3) µm, (51.4 ± 2.6) µm, (-5.7 ± 1.8) µm, and (54.7 ± 2.2) µm. Among the epithelial pachymetry parameters, Min-Max and Min showed high diagnostic efficiency for keratoconus (AUC = 0.967, 0.968; Youden index = 0.882, 0.829); inferior, I-S, and central parameters showed moderate efficiency (AUC = 0.794, 0.728, 0.715; Youden index = 0.461, 0.564, 0.565). The IT-SN, I-S, Min, Min-median, Min-Max, and central corneal thickness averages were (-82.8 ± 65.2) µm, (-72.5 ± 55.6) µm, (401.2 ± 83.8) µm, (-83.4 ± 48.2) µm, (-169.7 ± 99.4) µm, and (453.4 ± 58.6) µm in keratoconic eyes.In normal eyes, these parameters were respectively (-23.7 ± 18.5) µm, (-20.5 ± 19.8) µm, (525.4 ± 28.1) µm, (-23.1 ± 4.8) µm, (-59.2 ± 23.3) µm, and (533.4 ± 28.8) µm. Min, Min-median, and Min-Max had preferable diagnostic efficiency (AUC = 0.952, 0.938, 0.924; Youden index = 0.784, 0.829, 0.721); central, IT-SN, and I-S parameters showed moderate efficiency (AUC = 0.891, 0.877, 0.862; Youden index = 0.643, 0.707, 0.580).

CONCLUSIONS: In observation with Fourier-domain OCT, keratoconus was characterized by central and inferior epithelial thinning, increased epithelial thickness variation, central cornea thinning, increase in asymmetry, and gaps between corneal IT-SN and I-S thickness.Fourier-domain OCT was confirmed to have high speed, high resolution, and noninvasion. The pachymetry parameters showed specific strengths in diagnosis of keratoconus, especially epithelial pachymetry parameters of Min and Min-Max, and corneal pachymetry parameters of Min, Min-median and Min-Max.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app