We have located links that may give you full text access.
EVALUATION STUDIES
JOURNAL ARTICLE
Cemented metal-on-metal total hip replacement with 28-mm head: prospective, long-term, clinical, radiological and metal ions data.
PURPOSE: The aim of this study was to evaluate the long-term clinical, radiological and metal ion blood concentration results following 28-mm metal-on-metal cemented total hip replacement using Metasul(®) acetabular component and polished, cannulated Allopro CF-30 (Sulzer-Medica, subsequently Centerpulse-Zimmer, Winterthur, Switzerland) femoral component.
METHODS: Prospective follow-up of patients operated between 1997 and 2000 at a district general hospital.
RESULTS: Seventy-nine patients (89 implants) with female predominance and median age of 66 years (IQR 45-87 years) were prospectively followed up for mean 13 years (11-14 years). There was significant improvement in Harris hip score (paired Student's t test p = 0.0001). The mean plasma cobalt and chromium levels were 1.3 µg/L (IQR 0.5-23.9) and 3.6 µg/L (IQR 1.0-22), respectively. Elevated plasma metal ions >7 µg/L were noted in four asymptomatic patients with negative ultrasound examination. Radiolucent lines were present in various zones but majority were stable. One femoral component was revised due to aseptic femoral loosening. Three acetabular and one femoral component have radiologically failed but not revised yet. The Kaplan-Meier survival at 14 years was 92 % for failure as endpoint.
CONCLUSION: The long-term survival of Metasul(®) cemented total hip replacements using 28-mm metal-on-metal head is comparable with metal on polyethylene bearing devices.
METHODS: Prospective follow-up of patients operated between 1997 and 2000 at a district general hospital.
RESULTS: Seventy-nine patients (89 implants) with female predominance and median age of 66 years (IQR 45-87 years) were prospectively followed up for mean 13 years (11-14 years). There was significant improvement in Harris hip score (paired Student's t test p = 0.0001). The mean plasma cobalt and chromium levels were 1.3 µg/L (IQR 0.5-23.9) and 3.6 µg/L (IQR 1.0-22), respectively. Elevated plasma metal ions >7 µg/L were noted in four asymptomatic patients with negative ultrasound examination. Radiolucent lines were present in various zones but majority were stable. One femoral component was revised due to aseptic femoral loosening. Three acetabular and one femoral component have radiologically failed but not revised yet. The Kaplan-Meier survival at 14 years was 92 % for failure as endpoint.
CONCLUSION: The long-term survival of Metasul(®) cemented total hip replacements using 28-mm metal-on-metal head is comparable with metal on polyethylene bearing devices.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app