Case Reports
Journal Article
Add like
Add dislike
Add to saved papers

Case report: paravalvular leak as a complication of percutaneous catheter ablation for atrial fibrillation.

An increasing number of reports reveal the safety and efficacy of catheter ablation for atrial fibrillation therapy even in patients with a history of prosthetic heart valve. This case report presents a 60 year old female patient who had a prosthetic mitral paravalvular leak as a complication of radiofrequency ablation. Surgery revealed a 15 mm hole and three broken knots in the posterolateral mitral annulus. This report demonstrates radiofrequency ablation for atrial fibrillation in patients with prosthetic mitral valve may cause a paravalvular leak likely as a result of thermal injury. Moreover, patient can be asymptomatic at first. The clinician should keep in mind such complication and the patient should be evaluated in terms of paravalvular leakage.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app