JOURNAL ARTICLE

Accuracy of a hand-held surgical navigation system for tibial resection in total knee arthroplasty

William D Bugbee, Arash Y Kermanshahi, Michelle M Munro, Julie C McCauley, Steven N Copp
Knee 2014, 21 (6): 1225-8
25311515

BACKGROUND: Accuracy of total knee arthroplasty (TKA) implant placement and overall limb are important goals of TKA technique.

METHODS: The accuracy and ease of use of an accelerometer-based hand-held navigation system for tibial resection during TKA was examined in 90 patients. Preoperative goals for sagittal alignment, navigation system assembly time, resection time, and tourniquet time were evaluated. Coronal and sagittal alignment was measured postoperatively.

RESULTS: The average coronal tibial component alignment was 0.43° valgus; 6.7% of patients had tibial coronal alignment outside of ±3° varus/valgus. The difference between the intraoperative goal and radiographically measured posterior tibial slope was 0.5°. The average time to completion of the tibial cut was 4.6 minutes.

CONCLUSION: The accelerometer-based hand-held navigation system was accurate for tibial coronal and sagittal alignment during TKA, with no additional surgical time compared with conventional instrumentation.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
25311515
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"