Validity and reliability of head posture measurement using Microsoft Kinect.
British Journal of Ophthalmology 2014 November
AIMS: To investigate the validity and reliability of Microsoft Kinect-based head tracker (KHT) for measuring head posture.
METHODS: Considering the cervical range of motion (CROM) as a reference, one-dimensional and three-dimensional (1D and 3D) head postures of 12 normal subjects (28-58 years of age; 6 women and 6 men) were obtained using the KHT. The KHT was validated by Pearson's correlation coefficient and intraclass correlation (ICC) coefficient. Test-retest reliability of the KHT was determined by its 95% limit of agreement (LoA) with the Bland-Altman plot. Face recognition success rate was evaluated for each head posture.
RESULTS: Measurements of 1D and 3D head posture performed using the KHT were very close to those of the CROM with correlation coefficients of 0.99 and 0.97 (p<0.05), respectively, as well as with an ICC of >0.99 and 0.98, respectively. The reliability tests of the KHT in terms of 1D and 3D head postures had 95% LoA angles of approximately ±2.5° and ±6.5°, respectively.
CONCLUSIONS: The KHT showed good agreement with the CROM and relatively favourable test-retest reliability. Considering its high performance, convenience and low cost, KHT could be clinically used as a head posture-measuring system.
METHODS: Considering the cervical range of motion (CROM) as a reference, one-dimensional and three-dimensional (1D and 3D) head postures of 12 normal subjects (28-58 years of age; 6 women and 6 men) were obtained using the KHT. The KHT was validated by Pearson's correlation coefficient and intraclass correlation (ICC) coefficient. Test-retest reliability of the KHT was determined by its 95% limit of agreement (LoA) with the Bland-Altman plot. Face recognition success rate was evaluated for each head posture.
RESULTS: Measurements of 1D and 3D head posture performed using the KHT were very close to those of the CROM with correlation coefficients of 0.99 and 0.97 (p<0.05), respectively, as well as with an ICC of >0.99 and 0.98, respectively. The reliability tests of the KHT in terms of 1D and 3D head postures had 95% LoA angles of approximately ±2.5° and ±6.5°, respectively.
CONCLUSIONS: The KHT showed good agreement with the CROM and relatively favourable test-retest reliability. Considering its high performance, convenience and low cost, KHT could be clinically used as a head posture-measuring system.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app