Accuracy of Individualized Custom Tibial Cutting Guides in UKA.
BACKGROUND: Component malposition is one of the major reasons for early failure of unicompartmental knee arthroplasty (UKA).
QUESTIONS/PURPOSES: It was investigated how reproducibly patient-specific instrumentation (PSI) achieved preoperatively planned placement of the tibial component in UKA specifically assessing coronal alignment, slope and flexion of the components and axial rotation.
PATIENTS AND METHODS: Based on computer tomography models of ten cadaver legs, PSI jigs were generated to guide cuts perpendicular to the tibial axis in the coronal and sagittal planes and in neutral axial rotation. Deviation ≥3° from the designed orientation in a postoperative CT was defined as outside the range of acceptable alignment.
RESULTS: Mean coronal alignment was 0.4 ± 3.2° varus with two outliers. Mean slope was 2.8 ± 3.9° with six components in excessive flexion. It was noted that the implants were put in a mean of 1.7 ± 8.0° of external rotation with seven outliers.
CONCLUSIONS: PSI helped achieve the planned coronal orientation of the component. The guides were less accurate in setting optimal tray rotation and slope.
QUESTIONS/PURPOSES: It was investigated how reproducibly patient-specific instrumentation (PSI) achieved preoperatively planned placement of the tibial component in UKA specifically assessing coronal alignment, slope and flexion of the components and axial rotation.
PATIENTS AND METHODS: Based on computer tomography models of ten cadaver legs, PSI jigs were generated to guide cuts perpendicular to the tibial axis in the coronal and sagittal planes and in neutral axial rotation. Deviation ≥3° from the designed orientation in a postoperative CT was defined as outside the range of acceptable alignment.
RESULTS: Mean coronal alignment was 0.4 ± 3.2° varus with two outliers. Mean slope was 2.8 ± 3.9° with six components in excessive flexion. It was noted that the implants were put in a mean of 1.7 ± 8.0° of external rotation with seven outliers.
CONCLUSIONS: PSI helped achieve the planned coronal orientation of the component. The guides were less accurate in setting optimal tray rotation and slope.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app