Temporal B0 field variation effects on MRSI of the human prostate at 7 T and feasibility of correction using an internal field probe.

NMR in Biomedicine 2014 November
Spectral degradations as a result of temporal field variations are observed in MRSI of the human prostate. Moving organs generate substantial temporal and spatial field fluctuations as a result of susceptibility mismatch with the surrounding tissue (i.e. periodic breathing, cardiac motion or random bowel motion). Nine patients with prostate cancer were scanned with an endorectal coil (ERC) on a 7-T MR scanner. Temporal B0 field variations were observed with fast dynamic B0 mapping in these patients. Simulations of dynamic B0 corrections were performed using zero- to second-order shim terms. In addition, the temporal B0 variations were applied to simulated MR spectra causing, on average, 15% underestimation of the choline/citrate ratio. Linewidth distortions and frequency shifts (up to 30 and 8 Hz, respectively) were observed. To demonstrate the concept of observing local field fluctuations in real time during MRSI data acquisition, a field probe (FP) tuned and matched for the (19)  F frequency was incorporated into the housing of the ERC. The data acquired with the FP were compared with the B0 field map data and used to correct the MRSI datasets retrospectively. The dynamic B0 mapping data showed variations of up to 30 Hz (0.1 ppm) over 72 s at 7 T. The simulated zero-order corrections, calculated as the root mean square, reduced the standard deviation (SD) of the dynamic variations by an average of 41%. When using second-order corrections, the reduction in the SD was, on average, 56%. The FP data showed the same variation range as the dynamic B0 data and the variation patterns corresponded. After retrospective correction, the MRSI data showed artifact reduction and improved spectral resolution. B0 variations can degrade the MRSI substantially. The simple incorporation of an FP into an ERC can improve prostate cancer MRSI without prior knowledge of the origin of the dynamic field distortions.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app