JOURNAL ARTICLE

Oesophageal Doppler and calibrated pulse contour analysis are not interchangeable within a goal-directed haemodynamic algorithm in major gynaecological surgery

A Feldheiser, O Hunsicker, H Krebbel, K Weimann, L Kaufner, K-D Wernecke, C Spies
British Journal of Anaesthesia 2014, 113 (5): 822-31
25107544

BACKGROUND: Evidence for the benefit of an intraoperative use of a goal-directed haemodynamic management has grown. We compared the oesophageal Doppler monitor (ODM, CardioQ-ODM™) with a calibrated pulse contour analysis (PCA, PiCCO2™) with regard to assessment of stroke volume (SV) changes after volume administration within a goal-directed haemodynamic algorithm during non-cardiac surgery.

METHODS: The data were obtained prospectively in patients with metastatic ovarian carcinoma undergoing cytoreductive surgery. During surgery, fluid challenges were performed as indicated by the goal-directed haemodynamic algorithm guided by the ODM. Monitors were compared regarding precision and trending. Clinical characteristics associated with trending were studied by extended regression analysis.

RESULTS: A total of 762 fluid challenges were performed in 41 patients resulting in 1524 paired measurements. The precision of ODM and PCA was 5.7% and 6.0% (P=0.80), respectively. Polar plot analysis revealed a poor trending between ODM and PCA with an angular bias of -7.1°, radial limits of agreement of -58.1° to 43.8°, and an angular concordance rate of 67.8%. Dose of norepinephrine (NE) (scaled 0.1 µg kg(-1) min(-1)) [adjusted odds ratio (OR) 0.606 (95% confidence interval, CI: 0.404-0.910); P=0.016] and changes in mean arterial pressure (MAP) to a fluid challenge (scaled 10%) [adjusted OR 0.733 (95% CI: 0.635-0.845); P<0.001] were associated with trending between ODM and PCA, whereas there was no relation to type of i.v. solution.

CONCLUSIONS: Despite a similar precision, ODM and PCA were not interchangeable with regard to measuring SV changes within a goal-directed haemodynamic algorithm. A decrease in interchangeability coincided with increasing NE levels and greater changes of MAP to a fluid challenge.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
25107544
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"