The randomized crossover comparison of airway sealing with the laryngeal mask airway Supreme(™) at three different intracuff pressures in children.
Paediatric Anaesthesia 2014 October
BACKGROUND: An intracuff pressure of 80 cm H2 O in the adult-sized laryngeal mask airway Supreme has been recommended to obtain a higher oropharyngeal leak pressure (OLP). However, the intracuff pressure for the higher OLP in the pediatric laryngeal mask airway Supreme could be different from that in the adult-sized laryngeal mask airway Supreme. Thus, we measured and compared OLP at three intracuff pressures of 40, 60, and 80 cm H2 O in the pediatric laryngeal mask airway Supreme.
METHODS: This study was designed as a randomized crossover study and enrolled 36 children, aged 0-108 months and 5-30 kg in weight. After insertion of the laryngeal mask airway Supreme, oropharyngeal leak pressure (OLP) was measured at intracuff pressures of 40, 60, and 80 cm H2 O according to one of six sequences produced on the basis of 3 × 6 Williams crossover design. During the intraoperative period, the laryngeal mask airway Supreme was maintained using the last intracuff pressure of the allocated sequence. The intraoperative and postoperative complications were compared among three maintenance intracuff pressures.
RESULTS: OLP at the intracuff pressure of 60 cm H2 O was significantly higher than that of 40 cm H2 O (17.9 ± 3.9 vs 16.9 ± 4.2 cm H2 O, P = 0.004) and was comparable with that of 80 cm H2 O (17.9 ± 3.9 vs 17.8 ± 4.6 cm H2 O, P = 0.938). There were no significant differences of intraoperative and postoperative complications among the three maintenance intracuff pressures.
CONCLUSION: Our results suggest that the use of an intracuff pressure of 60 cm H2 O in pediatric laryngeal mask airway Supreme provides a higher OLP compared with 40 cm H2 O.
METHODS: This study was designed as a randomized crossover study and enrolled 36 children, aged 0-108 months and 5-30 kg in weight. After insertion of the laryngeal mask airway Supreme, oropharyngeal leak pressure (OLP) was measured at intracuff pressures of 40, 60, and 80 cm H2 O according to one of six sequences produced on the basis of 3 × 6 Williams crossover design. During the intraoperative period, the laryngeal mask airway Supreme was maintained using the last intracuff pressure of the allocated sequence. The intraoperative and postoperative complications were compared among three maintenance intracuff pressures.
RESULTS: OLP at the intracuff pressure of 60 cm H2 O was significantly higher than that of 40 cm H2 O (17.9 ± 3.9 vs 16.9 ± 4.2 cm H2 O, P = 0.004) and was comparable with that of 80 cm H2 O (17.9 ± 3.9 vs 17.8 ± 4.6 cm H2 O, P = 0.938). There were no significant differences of intraoperative and postoperative complications among the three maintenance intracuff pressures.
CONCLUSION: Our results suggest that the use of an intracuff pressure of 60 cm H2 O in pediatric laryngeal mask airway Supreme provides a higher OLP compared with 40 cm H2 O.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app