Next generation coronary CT angiography: in vitro evaluation of 27 coronary stents.
European Radiology 2014 November
OBJECTIVES: To evaluate in-stent lumen visibility of 27 modern and commonly used coronary stents (16 individual stent types, two stents at six different sizes each) utilising a third-generation dual-source CT system.
METHODS: Stents were implanted in a plastic tube filled with contrast. Examinations were performed parallel to the system's z-axis for all stents (i.e. 0°) and in an orientation of 90° for stents with a diameter of 3.0 mm. Two stents were evaluated in different diameters (2.25 to 4.0 mm). Examinations were acquired with a collimation of 96 × 0.6 mm, tube voltage of 120 kVp with 340 mAs tube current. Evaluation was performed using a medium-soft (Bv40), a medium-sharp (Bv49) and a sharp (Bv59) convolution kernel optimised for vascular imaging.
RESULTS: Mean visible stent lumen of stents with 3.0 mm diameter ranged from 53.3 % (IQR 48.9 - 56.7 %) to 73.9 % (66.7 - 76.7 %), depending on the kernel used at 0°, and was highest at an orientation of 90° with 80.0 % (75.6 - 82.8 %) using the Bv59 kernel, strength 4. Visible stent lumen declined with decreasing stent size.
CONCLUSIONS: Use of third-generation dual-source CT enables stent lumen visibility of up to 80 % in metal stents and 100 % in bioresorbable stents.
KEY POINTS: • Blooming artefacts impair in-stent lumen visibility of coronary stents in CT angiography. • CT enables stent lumen visibility of up to 80 % in metal stents. • Stent lumen visibility varies with stent orientation and size. • CT angiography may be a valid alternative for detecting in-stent restenosis.
METHODS: Stents were implanted in a plastic tube filled with contrast. Examinations were performed parallel to the system's z-axis for all stents (i.e. 0°) and in an orientation of 90° for stents with a diameter of 3.0 mm. Two stents were evaluated in different diameters (2.25 to 4.0 mm). Examinations were acquired with a collimation of 96 × 0.6 mm, tube voltage of 120 kVp with 340 mAs tube current. Evaluation was performed using a medium-soft (Bv40), a medium-sharp (Bv49) and a sharp (Bv59) convolution kernel optimised for vascular imaging.
RESULTS: Mean visible stent lumen of stents with 3.0 mm diameter ranged from 53.3 % (IQR 48.9 - 56.7 %) to 73.9 % (66.7 - 76.7 %), depending on the kernel used at 0°, and was highest at an orientation of 90° with 80.0 % (75.6 - 82.8 %) using the Bv59 kernel, strength 4. Visible stent lumen declined with decreasing stent size.
CONCLUSIONS: Use of third-generation dual-source CT enables stent lumen visibility of up to 80 % in metal stents and 100 % in bioresorbable stents.
KEY POINTS: • Blooming artefacts impair in-stent lumen visibility of coronary stents in CT angiography. • CT enables stent lumen visibility of up to 80 % in metal stents. • Stent lumen visibility varies with stent orientation and size. • CT angiography may be a valid alternative for detecting in-stent restenosis.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app