Transcatheter closure of perimembranous ventricular septal defects with ductal occluders.
Cardiology in the Young 2015 June
BACKGROUND: To study the feasibility and complications associated with the use of ductal occluders for closure of perimembranous ventricular septal defects.
METHODS: A total of 126 patients, ranging from 1 to 41 years of age (median - 8 years), underwent closure of ventricular septal defects from August 2010 to April 2013. Small- and moderate-sized defects were closed using first-generation Patent ductus arteriosus occluders or Amplatzer Duct Occluder-II. Patients were followed up for the development of complications such as heart block, aortic regurgitation, and tricuspid regurgitation.
RESULTS: Patent ductus arteriosus occluders were used in 81 patients, and the Amplatzer Duct Occluder-II device in 45 patients. The devices were successfully deployed in 99.2% of the cases. One patient had embolisation of an Amplatzer Duct Occluder-II device soon after deployment. There was one case of transient complete heart block (0.8%) needing temporary pacing, and two cases of isoarrhythmic atrioventricular dissociation (1.6%). One patient developed late-onset complete heart block 15 months after the procedure and underwent permanent pacemaker implantation. There were no instances of new-onset aortic regurgitation. New-onset mild tricuspid regurgitation was seen in two patients. Of the patients, three had small residual shunts on follow-up, without haemolysis.
CONCLUSIONS: Duct occluders can be used to effectively close small- and moderate-sized ventricular septal defects. The incidence of complete heart block and valvular regurgitations are much less than reported with other devices, and they are cost-effective.
METHODS: A total of 126 patients, ranging from 1 to 41 years of age (median - 8 years), underwent closure of ventricular septal defects from August 2010 to April 2013. Small- and moderate-sized defects were closed using first-generation Patent ductus arteriosus occluders or Amplatzer Duct Occluder-II. Patients were followed up for the development of complications such as heart block, aortic regurgitation, and tricuspid regurgitation.
RESULTS: Patent ductus arteriosus occluders were used in 81 patients, and the Amplatzer Duct Occluder-II device in 45 patients. The devices were successfully deployed in 99.2% of the cases. One patient had embolisation of an Amplatzer Duct Occluder-II device soon after deployment. There was one case of transient complete heart block (0.8%) needing temporary pacing, and two cases of isoarrhythmic atrioventricular dissociation (1.6%). One patient developed late-onset complete heart block 15 months after the procedure and underwent permanent pacemaker implantation. There were no instances of new-onset aortic regurgitation. New-onset mild tricuspid regurgitation was seen in two patients. Of the patients, three had small residual shunts on follow-up, without haemolysis.
CONCLUSIONS: Duct occluders can be used to effectively close small- and moderate-sized ventricular septal defects. The incidence of complete heart block and valvular regurgitations are much less than reported with other devices, and they are cost-effective.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app