Low-flow oxygen for positive pressure ventilation of preterm infants in the delivery room.

BACKGROUND: The recent newborn resuscitation guidelines have recommended that a pulse oximeter and oxygen blender be used to keep oxygen saturation (SpO2) within the target range. However, an oxygen blender and compressed air are not generally available in delivery rooms.

OBJECTIVES: To determine whether using low-flow oxygen at 0.5-1 liters/min for positive pressure ventilation (PPV) via a self-inflating bag (SIB) without a reservoir is effective and able to maintain SpO2 within the target range.

METHODS: Infants with a gestational age (GA) ≤32 weeks who initially required PPV after birth were enrolled. PPV was performed with low-flow oxygen at 0.5-1 liters/min via an SIB without a reservoir, and the flow was adjusted in a stepwise manner (from 0.5 to 0.8 to 1 liters/min) to keep SpO2 in the target range. If the heart rate was still <100/min or SpO2 was <70% at 3 min or chest compression was needed, then 100% oxygen was provided.

RESULTS: Forty-seven infants were enrolled in the study with a median (interquartile range) GA and birth weight of 28 (27-30) weeks and 1,060 (770-1,360) g, respectively. Twelve infants were initially intubated and switched to 100% oxygen (n = 12) due to ineffective ventilation, which occurred predominately in lower GA infants with intrapartum fetal distress. Thirty infants were successfully resuscitated with low-flow oxygen PPV (success rate 85.7%, 30/35), and >80% of their SpO2 distribution during PPV was between the 3rd and 97th percentiles of the reference range.

CONCLUSION: Low-flow oxygen for PPV via an SIB used in this study should be sufficient for providing oxygen in resuscitation of preterm infants as long as adequate ventilation is evident. This technique is simple and could be useful in a resource-limited setting.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app