COMPARATIVE STUDY
JOURNAL ARTICLE

All-inside meniscal repair devices compared with their matched inside-out vertical mattress suture repair: introducing 10,000 and 100,000 loading cycles

Claudio Rosso, Sebastian Müller, Daniel M Buckland, Tanja Schwenk, Simon Zimmermann, Michael de Wild, Victor Valderrabano
American Journal of Sports Medicine 2014, 42 (9): 2226-33
24966304

BACKGROUND: All-inside arthroscopic meniscal repairs are favored by most clinicians because of their lower complication rate and decreased morbidity compared with inside-out techniques. Until now, only 1000 cycles have been used for biomechanical testing.

HYPOTHESIS: All-inside meniscal repairs will show inferior biomechanical response to cyclic loading (up to 100,000 cycles) and load-to-failure testing compared with inside-out suture controls.

STUDY DESIGN: Controlled laboratory study.

METHODS: Bucket-handle tears in 72 porcine menisci were repaired using the Omnispan and Fast-Fix 360 (all-inside devices) and Orthocord 2-0 and Ultrabraid 2-0 sutures (matched controls). Initial displacement, displacement after cyclic loading (100, 500, 1000, 2000, 5000, 10,000, and 100,000 cycles) between 5 and 20 N, ultimate load to failure, and mode of failure were recorded, as well as stiffness.

RESULTS: Initial displacement and displacement after cyclic loading were not different between the groups. The Omnispan repair demonstrated the highest load-to-failure force (mean ± SD, 151.3 ± 21.5 N) and was significantly stronger than all the other constructs (Orthocord 2-0, 105.5 ± 20.4 N; Ultrabraid 2-0, 93.4 ± 22.5 N; Fast-Fix 360, 76.6 ± 14.2 N) (P < .0001 for all). The Orthocord vertical inside-out mattress repair was significantly stronger than the Fast-Fix 360 repair (P = .003). The Omnispan (30.8 ± 3.5 N/mm) showed significantly higher stiffness compared with the Ultrabraid 2-0 (22.9 ± 6.9 N/mm, P < .0001) and Fast-Fix 360 (23.7 ± 3.9 N/mm, P = .001). The predominant mode of failure was suture failure.

CONCLUSION: All-inside meniscal devices show comparable biomechanical properties compared with inside-out suture repair in cyclic loading, even after 100,000 cycles.

CLINICAL RELEVANCE: Eight to 10 weeks of rehabilitation might not pose a problem for all repairs in this worst-case scenario.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
24966304
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"