AIM: Providing consistent levels of oxygen saturation (SpO2 ) for infants in neonatal intensive care units is not easy. This study explored how effectively the Auto-Mixer(®) algorithm automatically adjusted fraction of inspired oxygen (FiO2 ) levels to maintain SpO2 within an intended range in extremely low birth weight infants receiving supplemental oxygen without mechanical ventilation.
METHODS: Twenty extremely low birth weight infants were randomly assigned to the Auto-Mixer(®) group or the manual intervention group and studied for 12 h. The SpO2 target was 85-93%, and the outcomes were the percentage of time SpO2 was within target, SpO2 variability, SpO2 >95%, oxygen received and manual interventions.
RESULTS: The percentage of time within intended SpO2 was 58 ± 4% in the Auto-Mixer(®) group and 33.7 ± 4.7% in the manual group, SpO2 >95% was 26.5% vs 54.8%, average SpO2 and FiO2 were 89.8% vs 92.2% and 37% vs 44.1%, and manual interventions were 0 vs 80 (p < 0.05). Brief periods of SpO2 < 85% occurred more frequently in the Auto-Mixer(®) group.
CONCLUSION: The Auto-Mixer(®) effectively increased the percentage of time that SpO2 was within the intended target range and decreased the time with high SpO2 in spontaneously breathing extremely low birth weight infants receiving supplemental oxygen.
METHODS: Twenty extremely low birth weight infants were randomly assigned to the Auto-Mixer(®) group or the manual intervention group and studied for 12 h. The SpO2 target was 85-93%, and the outcomes were the percentage of time SpO2 was within target, SpO2 variability, SpO2 >95%, oxygen received and manual interventions.
RESULTS: The percentage of time within intended SpO2 was 58 ± 4% in the Auto-Mixer(®) group and 33.7 ± 4.7% in the manual group, SpO2 >95% was 26.5% vs 54.8%, average SpO2 and FiO2 were 89.8% vs 92.2% and 37% vs 44.1%, and manual interventions were 0 vs 80 (p < 0.05). Brief periods of SpO2 < 85% occurred more frequently in the Auto-Mixer(®) group.
CONCLUSION: The Auto-Mixer(®) effectively increased the percentage of time that SpO2 was within the intended target range and decreased the time with high SpO2 in spontaneously breathing extremely low birth weight infants receiving supplemental oxygen.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app