COMPARATIVE STUDY
JOURNAL ARTICLE
RANDOMIZED CONTROLLED TRIAL
RESEARCH SUPPORT, NON-U.S. GOV'T
Add like
Add dislike
Add to saved papers

Comparison study of upper arm and forearm non-invasive blood pressures in adult Emergency Department patients.

BACKGROUND: Forearm blood pressures have been suggested as an alternative site to measure blood pressures when the upper arm is unavailable. However there is little evidence utilising clinical populations to support this substitution.

OBJECTIVES: To determine agreement between blood pressures measured in the left upper arm and forearm using a singular oscillometric non-invasive device in adult Emergency Department patients. The secondary objective was to explore the relationship of blood pressure differences with age, sex, ethnicity, smoking history and obesity.

DESIGN: Single centre comparison study.

SETTING: Adult Emergency Department, Tertiary Trauma Centre.

PARTICIPANTS: Forty-four participants who met inclusion/exclusion criteria selected sequentially from the Emergency Department arrival board.

METHODS: A random assignment of order of measurement for left upper arm and forearm blood pressures was utilised. Participants were eligible if they were aged 18 years or older, had been assigned an Australasian Triage Scale code of 2, 3, 4, or 5, were able to consent, and able to have blood pressures measured on their left arm whilst lying at a 45° angle. The Bland-Altman method of statistical analysis was used, with the level of agreement for clinical acceptability for the systolic, diastolic and mean arterial pressure defined as ±10 mmHg.

RESULTS: The forearm measure overestimated systolic (mean difference 2.2 mmHg, 95% limits of agreement ±19 mmHg), diastolic (mean difference 3.4 mmHg, 95% limits of agreement ±14.4 mmHg), and mean arterial pressures (mean difference 4.1 mmHg, 95% limits of agreement ±13.7 mmHg). The systolic measure was not significantly different from zero. Evidence of better agreement was found with upper arm/forearm systolic measures below 140 mmHg compared to systolic measures above 140 mmHg using the Levene's test (p=0.002, F-statistic=11.09). Blood pressure disparity was not associated with participant characteristics.

CONCLUSIONS: Forearm measures cannot routinely replace upper arm measures for blood pressure measurement. If the clinical picture requires use of forearm blood pressure, the potential variance from an upper arm measure is ±19 mmHg for systolic pressure, although the variability may be close to ±10 mmHg if the systolic blood pressure is below 140 mmHg.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app