Analytical validation of whole exome and whole genome sequencing for clinical applications.

BACKGROUND: Whole exome and genome sequencing (WES/WGS) is now routinely offered as a clinical test by a growing number of laboratories. As part of the test design process each laboratory must determine the performance characteristics of the platform, test and informatics pipeline. This report documents one such characterization of WES/WGS.

METHODS: Whole exome and whole genome sequencing was performed on multiple technical replicates of five reference samples using the Illumina HiSeq 2000/2500. The sequencing data was processed with a GATK-based genome analysis pipeline to evaluate: intra-run, inter-run, inter-mode, inter-machine and inter-library consistency, concordance with orthogonal technologies (microarray, Sanger) and sensitivity and accuracy relative to known variant sets.

RESULTS: Concordance to high-density microarrays consistently exceeds 97% (and typically exceeds 99%) and concordance between sequencing replicates also exceeds 97%, with no observable differences between different flow cells, runs, machines or modes. Sensitivity relative to high-density microarray variants exceeds 95%. In a detailed study of a 129 kb region, sensitivity was lower with some validated single-base insertions and deletions "not called". Different variants are "not called" in each replicate: of all variants identified in WES data from the NA12878 reference sample 74% of indels and 89% of SNVs were called in all seven replicates, in NA12878 WGS 52% of indels and 88% of SNVs were called in all six replicates. Key sources of non-uniformity are variance in depth of coverage, artifactual variants resulting from repetitive regions and larger structural variants.

CONCLUSION: We report a comprehensive performance characterization of WES/WGS that will be relevant to offering laboratories, consumers of genome sequencing and others interested in the analytical validity of this technology.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app