Journal Article
Multicenter Study
Research Support, Non-U.S. Gov't
Add like
Add dislike
Add to saved papers

Postprocedural aortic regurgitation in balloon-expandable and self-expandable transcatheter aortic valve replacement procedures: analysis of predictors and impact on long-term mortality: insights from the FRANCE2 Registry.

Circulation 2014 April 2
BACKGROUND: Significant postprocedural aortic regurgitation (AR) is observed in 10% to 20% of cases after transcatheter aortic valve replacement (TAVR). The prognostic value and the predictors of such a complication in balloon-expandable (BE) and self-expandable (SE) TAVR remain unclear.

METHODS AND RESULTS: TAVR was performed in 3195 consecutive patients at 34 hospitals. Postprocedural transthoracic echocardiography was performed in 2769 (92%) patients of the eligible population, and these patients constituted the study group. Median follow-up was 306 days (Q1-Q3=178-490). BE and SE devices were implanted in 67.6% (n=1872) and 32.4% (n=897). Delivery was femoral (75.3%) or nonfemoral (24.7%). A postprocedural AR≥grade 2 was observed in 15.8% and was more frequent in SE (21.5%) than in BE-TAVR (13.0%, P=0.0001). Extensive multivariable analysis confirmed that the use of a SE device was one of the most powerful independent predictors of postprocedural AR≥grade 2. For BE-TAVR, 8 independent predictors of postprocedural AR≥grade 2 were identified including femoral delivery (P=0.04), larger aortic annulus (P=0.0004), and smaller prosthesis diameter (P=0.0001). For SE-TAVR, 2 independent predictors were identified including femoral delivery(P=0.0001). Aortic annulus and prosthesis diameter were not predictors of postprocedural AR for SE-TAVR. A postprocedural AR≥grade 2, but not a postprocedural AR=grade 1, was a strong independent predictor of 1-year mortality for BE (hazard ratio=2.50; P=0.0001) and SE-TAVR (hazard ratio=2.11; P=0.0001). Although postprocedural AR≥grade 2 was well tolerated in patients with AR≥grade 2 at baseline (1-year mortality=7%), it was associated with a very high mortality in other subgroups: renal failure (43%), AR
CONCLUSIONS: Postprocedural AR≥grade 2 was observed in 15.8% of successful TAVR and was the strongest independent predictor of 1-year mortality. The use of the SE device was a powerful independent predictor of postprocedural AR≥grade 2.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app