Development and validation of RP-HPLC method with ultraviolet detection for estimation of montelukast in rabbit plasma: Application to preclinical pharmacokinetics.
Journal of Young Pharmacists : JYP 2013 December
OBJECTIVE: To develop a liquid-liquid extraction based reverse phase liquid chromatography method for estimation of montelukast in rabbit plasma.
METHODS: Chromatographic separation was carried out using Phenomenex Luna C18 column (250 mm × 4.6 mm × 5 μm) with mobile phase composed of ammonium acetate buffer (20 Mm), pH 5.5 and acetonitrile in 20:80, v/v ratio. The analyte was monitored with UV detector at 345 nm. The developed method was validated with respect to linearity, accuracy, precision, specificity and stability.
RESULTS: The peak area ratio of montelukast (MKS) to that of internal standard was used for the quantification of samples. Calibration curves were linear in the concentration range of 20-2000 ng mL(-1). The LOD and LLOQ of present method were found out to be 10 ng mL(-1) and 20 ng mL(-1) respectively. The intra-day and inter-day %CV values for MKS were below 6.06% and 8.43%. Intra-day and inter-day accuracies were within 95.81% and 110.90%, respectively. Extraction recoveries of drug from rabbit plasma were >66.47%.
CONCLUSION: A simple, alternative, reproducible and sensitive HPLC-UV method was developed for MKS that can be used in preclinical pharmacokinetics.
METHODS: Chromatographic separation was carried out using Phenomenex Luna C18 column (250 mm × 4.6 mm × 5 μm) with mobile phase composed of ammonium acetate buffer (20 Mm), pH 5.5 and acetonitrile in 20:80, v/v ratio. The analyte was monitored with UV detector at 345 nm. The developed method was validated with respect to linearity, accuracy, precision, specificity and stability.
RESULTS: The peak area ratio of montelukast (MKS) to that of internal standard was used for the quantification of samples. Calibration curves were linear in the concentration range of 20-2000 ng mL(-1). The LOD and LLOQ of present method were found out to be 10 ng mL(-1) and 20 ng mL(-1) respectively. The intra-day and inter-day %CV values for MKS were below 6.06% and 8.43%. Intra-day and inter-day accuracies were within 95.81% and 110.90%, respectively. Extraction recoveries of drug from rabbit plasma were >66.47%.
CONCLUSION: A simple, alternative, reproducible and sensitive HPLC-UV method was developed for MKS that can be used in preclinical pharmacokinetics.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app