COMPARATIVE STUDY
JOURNAL ARTICLE

Femoral and tibial torsion measurement in children and adolescents: comparison of 3D models based on low-dose biplanar radiography and low-dose CT

Andrea B Rosskopf, Leonhard E Ramseier, Reto Sutter, Christian W A Pfirrmann, Florian M Buck
AJR. American Journal of Roentgenology 2014, 202 (3): W285-91
24555627

OBJECTIVE: The purpose of this study was to evaluate the interchangeability and reliability of femoral and tibial torsion measurements in children using 3D models based on biplanar radiography compared with CT measurements.

MATERIALS AND METHODS: Femoral and tibial torsion were measured in 50 patients (mean age, 10.9 years; range, 4.7-14.8 years) using 3D models based on low-dose biplanar radiography by two independent readers. Measurements on transverse CT images by two independent readers served as the reference standard. Intermethod and interreader agreement was calculated using descriptive statistics, intraclass correlation coefficient (ICC), and Bland-Altman analysis.

RESULTS: Femoral and tibial torsion were -6°-65° and 6°-51° for 3D models based on biplanar radiography and -13°-59° and 4°-52° for CT measurements. The average difference (±SD) between the two methods was 4.9°±3.8° and 5.5°±4.1°, respectively. The intermethod ICC for biplanar radiography was 0.90 (95% CI, 0.87-0.92) for femoral torsion and 0.75 (0.68-0.80) for tibial torsion. The interreader ICC was 0.93-0.97. Mean measurement differences between the two biplanar radiography readers were 3.4° (0.0°-11.0°) for femoral torsion and 3.9° (0.0°-15.0°) for tibial torsion. Mean interreader differences at CT were 3.3° (0.0°-9.0°) for femoral and 3.0° (0.0°-10.0°) for tibial torsion. There was no trend for larger intermethod differences with decreasing age of the children.

CONCLUSION: Femoral and tibial torsion measurements in children using 3D models based on biplanar radiography are comparable to CT measurement results. Despite skeletal immaturity, torsion measurements in children on biplanar radiography seem to be as reliable as those on CT images.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
24555627
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"