Iliosacral screw insertion using CT-3D-fluoroscopy matching navigation

Masaki Takao, Takashi Nishii, Takashi Sakai, Hideki Yoshikawa, Nobuhiko Sugano
Injury 2014, 45 (6): 988-94
24507831

BACKGROUND: Percutaneous iliosacral screw insertion requires substantial experience and detailed anatomical knowledge to find the proper entry point and trajectory even with the use of a navigation system. Our hypothesis was that three-dimensional (3D) fluoroscopic navigation combined with a preoperative computed tomography (CT)-based plan could enable surgeons to perform safe and reliable iliosacral screw insertion. The purpose of the current study is two-fold: (1) to demonstrate the navigation accuracy for sacral fractures and sacroiliac dislocations on widely displaced cadaveric pelves; and (2) to report the technical and clinical aspects of percutaneous iliosacral screw insertion using the CT-3D-fluoroscopy matching navigation system.

METHODS: We simulated three types of posterior pelvic ring disruptions with vertical displacements of 0, 1, 2 and 3cm using cadaveric pelvic rings. A total of six fiducial markers were fixed to the anterior surface of the sacrum. Target registration error over the sacrum was assessed with the fluoroscopic imaging centre on the second sacral vertebral body. Six patients with pelvic ring fractures underwent percutaneous iliosacral screw placement using the CT-3D-fluoroscopy matching navigation. Three pelvic ring fractures were classified as type B2 and three were classified as type C1 according to the AO-OTA classification. Iliosacral screws for the S1 and S2 vertebra were inserted.

RESULTS: The mean target registration error over the sacrum was 1.2mm (0.5-1.9mm) in the experimental study. Fracture type and amount of vertical displacement did not affect the target registration error. All 12 screws were positioned correctly in the clinical series. There were no postoperative complications including nerve palsy. The mean deviation between the planned and the inserted screw position was 2.5mm at the screw entry point, 1.8mm at the area around the nerve root tunnels and 2.2mm at the tip of the screw.

CONCLUSION: The CT-3D-fluoroscopy matching navigation system was accurate and robust regardless of pelvic ring fracture type and fragment displacement. Percutaneous iliosacral screw insertion with the navigation system is clinically feasible.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
24507831
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"