Evaluation of a liquid-phase colorimetric method for rapid antibacterial susceptibility testing.

BACKGROUND AND OBJECTIVE: Early determination of antibacterial susceptibility increases the success of therapy, decreases unnecessary use of antibacterial agents and side-effects, and lowers the overall healthcare cost.

MATERIALS AND METHODS: A rapid colorimetric method for antibacterial susceptibility testing named Qui-Sensitest (Inovative Biotechnology Organization, Istanbul, Turkey) was evaluated in this study.

RESULTS: Qui-Sensitest proved to be a reliable rapid test for determining antibacterial susceptibility having an overall agreement of 97.6% with Kirby Bauer disk diffusion test results for enteric bacteria with 0.4% of major discrepancies and 2.0% of minor discrepancy.

CONCLUSION: Since the test makes the results available in 3.5-6 hours, it can provide the means for choosing the right treatment regimen the same day the infectious agent is isolated.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app