Seventeen-millimeter St. Jude Medical Regent valve in patients with small aortic annulus: dose moderate prosthesis-patient mismatch matter?

BACKGROUND: The study was designed to evaluate the effects of moderate prosthesis-patient mismatch (defined as 0.65 cm(2)/m(2) <indexed effective orifice area ≤ 0.85 cm(2)/m(2)) on midterm outcomes after isolated aortic valve replacement with a 17-mm St. Jude Medical Regent valve in a large series of patients, and to determine if these effects are influenced by patient confounding variables.

METHODS: One-hundred and six patients with and without moderate prosthesis-patient mismatch early after implantation of a 17-mm Regent valve at aortic position were included. Both clinical and echocardiographic assessments were performed preoperatively, at discharge and during follow-up period (mean follow-up time 52.6 ± 11.9 months).

RESULTS: The prevalence of moderate prosthesis-patient mismatch was documented in 46 patients (43.4%) at discharge. During the follow-up period, no difference in the regression of left ventricular mass, decrease of transvalvular pressure gradients, mortality and prosthesis-related complications was observed between patients with and without moderate prosthesis-patient mismatch. After adjustment for several risk factors, moderate prosthesis-patient mismatch was associated with increased midterm mortality in patients with baseline left ventricular ejection fraction<50% (HR: 1.80, p=0.02), but with normal prognosis in those with preserved LV function. Younger age (cut off value=65 years) was not an independent predictor of increased midterm mortality and valve-related complications in patients with moderate prosthesis-patient mismatch.

CONCLUSIONS: Moderate prosthesis-patient mismatch after aortic valve replacement with a small mechanical prosthesis is associated with increased mortality and adverse events in patients with pre-existing left ventricular dysfunction. Selected patients with small aortic annulus can experience satisfactory clinical improvements and midterm survival after aortic valve replacement with a 17-mm Regent valve.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app