Measuring blood pressure in pregnancy and postpartum: assessing the reliability of automated measuring devices.

OBJECTIVE: To determine the reliability of automated blood pressure (BP) measuring devices in hypertensive pregnancies.

METHODS: Three sequential measurements were taken using mercury sphygmomanometry and an automated device in pregnant hypertensive women (n = 89).

RESULTS: Systolic BP (SBP) was higher with the automated device (mean difference 2.5 mmHg, 95% CI: 1.9-3.2 mmHg), whereas diastolic BP (DBP) was higher with the mercury sphygmomanometer (mean difference 2.0 mmHg, 95% CI: 1.5-2.6 mmHg). Within-patient variability for the automated device was: SBP 4.8 mmHg (95% CI: 4.6-5.2 mmHg), DBP 3.4 mmHg (95% CI: 3.1-3.7 mmHg); and for mercury sphygmomanometry: SBP 3.7 mmHg (95% CI: 3.5-4.0 mmHg), DBP 3.0 mmHg (95% CI: 2.8-3.3 mmHg).

CONCLUSION: Outpatient automated BP measurement agrees well with mercury sphygmomanometry in hypertensive pregnancies.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app