JOURNAL ARTICLE

Metaphyseal screw augmentation of the LISS-PLT plate with polymethylmethacrylate improves angular stability in osteoporotic proximal third tibial fractures: a biomechanical study in human cadaveric tibiae

Michael Goetzen, Tomas Nicolino, Ladina Hofmann-Fliri, Michael Blauth, Markus Windolf
Journal of Orthopaedic Trauma 2014, 28 (5): 294-9
24296600

OBJECTIVES: The incidence of osteoporotic proximal tibial fractures has increased during the last 2 decades. A promising approach in osteoporotic fracture fixation is polymethylmethacrylate-based cement augmentation of implants to gain better implant purchase in the bone. This study investigates the biomechanical benefits of screw augmentation in less invasive stabilization system-proximal lateral tibial (LISS-PLT) plates in cadaveric extraarticular comminuted proximal tibial fractures (OTA-41-A3.3).

METHODS: Standardized extraarticular proximal tibial fractures were stabilized with the LISS-PLT plate in 6 paired osteoporotic cadaveric tibiae. Bone mineral density was measured with high-resolution, quantitative computed tomography scans to identify bone quality. In the augmented group, the 5 proximal screws of the LISS-PLT plate were augmented with 1 mL of bone cement each, whereas the contralateral tibia was instrumented conventionally as the control. Cyclic axial loading was applied to each specimen with a starting load of 150 N, using a ramp of 0.05 N per cycle to 10-mm axial displacement. Varus displacement was identified from anterior-posterior radiographs.

RESULTS: Bone mineral density showed no significant difference between the 2 groups (P = 0.47). The nonaugmented group reached 9417 load cycles (SD 753) until failure, compared with 14,792 load cycles (SD 2088) in the augmented group (P = 0.002). In the early-onset failure (deformation at 8250 load cycles), varus displacement was significantly smaller in the augmented group (0.46 degrees, SD 0.6) than in the nonaugmented group (3.23 degrees, SD 1.7) (P = 0.01).

CONCLUSIONS: This biomechanical study showed that cement augmentation of the LISS-PLT plate screws in osteoporotic proximal extraarticular tibial fractures significantly lowers the propensity toward screw migration and secondary varus displacement.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
24296600
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"