Validation of a non-invasive pulse CO-oximetry based hemoglobin estimation in normal blood donors.
Transfusion and Apheresis Science 2014 Februrary
INTRODUCTION: Non-invasive hemoglobin estimation may increase the recruitment of blood donors. CO-oximetry hemoglobin estimation is a non-invasive method used to estimate the hemoglobin level. The primary objective of this study is to validate the pulse CO-oximetry based hemoglobin estimation in normal blood donors.
METHODS: We conducted a prospective observational study on 106 in a tertiary care hospital blood bank over a period of 4 weeks. We performed a Spot Hemoglobin concentration (Sp Hb) using Masimo Pronto-7 Pulse CO-oximetry, and compared it to a venous sample Hb concentration (Reference Hemoglobin; Ref Hb) measured using Abbott CELL-DYN Sapphire hematology analyzer. Age, gender, weight, height, blood pressure and reference hemoglobin were used in the multivariable linear regression model of the difference in measurement.
RESULTS: Total of 106 donors (98 males, 8 females) were enrolled with a mean age and Ref Hb of 27 years (SD 6.2; 18-49) and 14.2 g/dL (SD 1.2; 11.5-17) respectively. The mean Sp Hb was 14.4 g/dL (SD 1.2;11.3-16.7). The mean difference between the Sp Hb and Ref Hb was 0.2 g/dL (SD 1.2;-4.5 to 3) with a correlation coefficient of 0.46 (R(2)=21%). In the multivariable model, height (p=0.015) and Hb level (p<0.001) were statistically significant predictors. A strong correlation was found between the two CO-oximetry Hb measurements (coefficient 0.78, R(2)=60%).
CONCLUSIONS: Our study validated the use of the CO-oximetry in blood donors. Larger prospective studies are needed to confirm our results.
METHODS: We conducted a prospective observational study on 106 in a tertiary care hospital blood bank over a period of 4 weeks. We performed a Spot Hemoglobin concentration (Sp Hb) using Masimo Pronto-7 Pulse CO-oximetry, and compared it to a venous sample Hb concentration (Reference Hemoglobin; Ref Hb) measured using Abbott CELL-DYN Sapphire hematology analyzer. Age, gender, weight, height, blood pressure and reference hemoglobin were used in the multivariable linear regression model of the difference in measurement.
RESULTS: Total of 106 donors (98 males, 8 females) were enrolled with a mean age and Ref Hb of 27 years (SD 6.2; 18-49) and 14.2 g/dL (SD 1.2; 11.5-17) respectively. The mean Sp Hb was 14.4 g/dL (SD 1.2;11.3-16.7). The mean difference between the Sp Hb and Ref Hb was 0.2 g/dL (SD 1.2;-4.5 to 3) with a correlation coefficient of 0.46 (R(2)=21%). In the multivariable model, height (p=0.015) and Hb level (p<0.001) were statistically significant predictors. A strong correlation was found between the two CO-oximetry Hb measurements (coefficient 0.78, R(2)=60%).
CONCLUSIONS: Our study validated the use of the CO-oximetry in blood donors. Larger prospective studies are needed to confirm our results.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app