Analysis of adverse events with Essure hysteroscopic sterilization reported to the Manufacturer and User Facility Device Experience database.
Journal of Minimally Invasive Gynecology 2013 November
STUDY OBJECTIVE: The Manufacturer and User Facility Device Experience database may be useful for clinicians using a Food and Drug Administration-approved medical device to identify the occurrence of adverse events and complications. We sought to analyze and investigate reports associated with the Essure hysteroscopic sterilization system (Conceptus Inc., Mountain View, CA) using this database.
DESIGN: Retrospective review of the Manufacturer and User Facility Device Experience database for events related to Essure hysteroscopic sterilization from November 2002 to February 2012 (Canadian Task Force Classification III).
SETTING: Online retrospective review.
PATIENTS: Online reports of patients who underwent Essure tubal sterilization.
INTERVENTION: Essure tubal sterilization.
MEASUREMENTS AND MAIN RESULTS: Four hundred fifty-seven adverse events were reported in the study period. Pain was the most frequently reported event (217 events [47.5%]) followed by delivery catheter malfunction (121 events [26.4%]). Poststerilization pregnancy was reported in 61 events (13.3%), of which 29 were ectopic pregnancies. Other reported events included perforation (90 events [19.7%]), abnormal bleeding (44 events [9.6%]), and microinsert malposition (33 events [7.2%]). The evaluation and management of these events resulted in an additional surgical procedure in 270 cases (59.1%), of which 44 were hysterectomies.
CONCLUSION: Sixty-one unintended poststerilization pregnancies were reported in the study period, of which 29 (47.5%) were ectopic gestations. Thus, ectopic pregnancy must be considered if a woman becomes pregnant after Essure hysteroscopic sterilization. Additionally, 44 women underwent hysterectomy after an adverse event reported to be associated with the use of the device.
DESIGN: Retrospective review of the Manufacturer and User Facility Device Experience database for events related to Essure hysteroscopic sterilization from November 2002 to February 2012 (Canadian Task Force Classification III).
SETTING: Online retrospective review.
PATIENTS: Online reports of patients who underwent Essure tubal sterilization.
INTERVENTION: Essure tubal sterilization.
MEASUREMENTS AND MAIN RESULTS: Four hundred fifty-seven adverse events were reported in the study period. Pain was the most frequently reported event (217 events [47.5%]) followed by delivery catheter malfunction (121 events [26.4%]). Poststerilization pregnancy was reported in 61 events (13.3%), of which 29 were ectopic pregnancies. Other reported events included perforation (90 events [19.7%]), abnormal bleeding (44 events [9.6%]), and microinsert malposition (33 events [7.2%]). The evaluation and management of these events resulted in an additional surgical procedure in 270 cases (59.1%), of which 44 were hysterectomies.
CONCLUSION: Sixty-one unintended poststerilization pregnancies were reported in the study period, of which 29 (47.5%) were ectopic gestations. Thus, ectopic pregnancy must be considered if a woman becomes pregnant after Essure hysteroscopic sterilization. Additionally, 44 women underwent hysterectomy after an adverse event reported to be associated with the use of the device.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app