Three-dimensional valve repair-the better care? Midterm results of a saddle-shaped, rigid annuloplasty ring in patients with ischemic mitral regurgitation

Ralf Guenzinger, Eike Philipp Schneider, Thomas Guenther, Petra Wolf, Domenico Mazzitelli, Ruediger Lange, Bernhard Voss
Journal of Thoracic and Cardiovascular Surgery 2014, 148 (1): 176-82
24176268

OBJECTIVES: Undersized ring annuloplasty is the treatment of choice for functional mitral regurgitation. However, recurrence of mitral regurgitation within the first years is frequent. The aim of this study was to analyze the functional and clinical outcome after mitral valve repair with the 3-dimensional saddle-shaped Edwards GeoForm (Edwards Lifesciences LLC, Irvine, Calif) annuloplasty ring in patients with ischemic mitral regurgitation.

METHODS: Between November 2006 and November 2012, 70 patients (mean age, 68 ± 10 years; mean left ventricular ejection fraction, 40% ± 15%) with functional mitral regurgitation due to ischemic cardiomyopathy underwent mitral valve repair with the Edwards GeoForm annuloplasty ring. Concomitant procedures, such as coronary artery bypass grafting (75.7%), tricuspid valve repair (25.7%), aortic valve replacement (8.6%), and the Maze procedure (4.3%), were performed in 92.9% of patients. Follow-up is 97% complete (mean, 3.0 ± 1.7 years). Transthoracic echocardiography was obtained 2.4 ± 1.7 years postoperatively.

RESULTS: Thirty-day mortality was 5.9%. Overall survival at 5 years was 71.3% ± 6.9%. At 4 years, overall freedom from recurrence of mitral regurgitation grade 3+ or greater was 92.5% ± 3.6%, and freedom from recurrence of mitral regurgitation grade 2+ or greater was 71.0% ± 8.7%. Three patients required a mitral valve-related reoperation for ring dehiscence. New York Heart Association functional class improved from 3.6 ± 0.6 to 1.6 ± 0.6 during follow-up (P < .05). Mean mitral valve pressure gradient was 3.3 ± 1.8 mm Hg across all ring sizes at the time of follow-up.

CONCLUSIONS: Mitral valve repair with the 3-dimensional saddle-shaped Edwards GeoForm annuloplasty ring in case of ischemic mitral regurgitation shows a low rate of recurrent regurgitation at 4 years. Clinically relevant mitral stenosis was not detected. The importance of secure anchoring of the device in the mitral annulus has to be emphasized to prevent ring dehiscence.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
24176268
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"