BACKGROUND/AIM: Estimating the measurement error (reliability) of three dimensional gait analysis (3DGA) is crucial to interpretation of gait data. The purpose of this study was to investigate the intra-rater reliability of 3DGA and apply a novel method of visualising reliability of gait cycle waveforms.

METHODS: A test re-test design was employed. A convenience sample of 30 healthy adults (18F; 12 M: mean age 30 ± 6.8 years) participated. Subjects walked along a 10 m walkway at their preferred gait speed and 3DGA data were collected using a VICON(®) 3DGA system. Testing was performed by the same investigator on two separate days within two weeks. Data from 10 trials were averaged and analysed.

ANALYSIS: The intraclass correlation coefficient (ICC), the standard error of measurement (SEM), minimal detectable change (MDC) and limits of agreement were calculated for kinetic and kinematic data. Bland and Altman plots were applied to gait cycle waveforms.

RESULTS: Spatio-temporal parameters such as cadence, step length, velocity, step time and step width were highly repeatable generating ICC's of 0.90 and above and low SEM/MDC. Range of joint movement across the gait cycle was generally more reliable than either minimum or maximum values and higher ICCs were obtained for movement in the sagittal plane. For kinematic data the standard error of measurement was low (≤ 5) for the majority of parameters. Transverse plane measurements showed poor reliability with lowest ICC's. ICCs for kinetic data ranged from 0.51 to 0.81.

CONCLUSIONS: Reliability of 3DGA has been estimated for our gait laboratory. Bland and Altman plots of gait cycle waveforms provide a useful addition to reliability analysis.

Comments

You need to log in or sign up for an account to be able to comment.

No comments yet, be the first to post one!

Related Papers

Available on the App Store

Available on the Play Store
Remove bar
Read by QxMD icon Read
24139682
×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"